目的:研究快速血糖检测仪与罗氏7600全自动生化分析仪检测血糖结果的相关性和差异,为临床治疗提供参考价值。方法应用全自动生化分析仪和快速血糖仪对200例糖尿病患者的静脉血和末梢血进行血糖检测,检测结果用SPSS13.0统计学软件进行分析。结果快速血糖仪测定血糖结果分别为(9.63士3.81)mmol/L、全自动生化分析仪测定结果为(10.89±3.23)mmol/L。二者结果呈正相关(r=0.972),差异有统计学意义(P<0.01);全自动生化分析仪测定的结果高于快速血糖仪,差异有统计学意义(P<0.05)。结论快速血糖分析仪具有检测快、操作方便优点,但其结果低于全自动生化分析仪。应定期与全自动生化分析仪做比对试验使测定结果有较高的准确性,为临床提供更准确的血糖检验结果。
Objective Discussion of fast blood glucose meter and Roche7600 automatic biochemical analyzer detection of glucose results of the correlation and dif erence,in order to provide reference for clinical treatment.Methods Application of the fast blood glucose meter and automatic biochemical analyzer in 200 cases of adult diabetes in peripheral blood and venous blood glucose testing,test results using SPSS13.0statistical software to analyze.Results The fast blood glucose meter and automatic biochemical analyzer for measuring blood glucose results respectively(9.63+3.81)mmol/L,(10.89+3.23)mmol/L.Two results were positively correlated(r=0.972),the dif erence was statistical y significant(<0.01);automatic biochemical analyzer determination results was higher than that of the fast blood glucose meter,the difference was statistical y significant(<0.05).Conclusion Rapid glucose analyzer has advantages of convenient operation,fast detection,but the result is lower than the automatic biochemical analyzer.Should be regular and automatic biochemistry analyzer compared with test the results have higher accuracy,for clinical provides more accurate glucose test results.
作者
王志强
WANG Zhi-qiang(Clinical Laboratory,Ordos Central Hospital,Ordos 017000,Inner Mongolia,China)
关键词
快速血糖仪
罗氏7600全自动生化分析仪
末梢全血血糖
静脉血糖
Fast blood glucose meter
Roche7600 automatic biochemical analyzer
Peripheral blood glucose
Venous blood glucose