钚量热计的研制 被引量:6

Development of Plutonium Calorimeter
下载PDF
导出
摘要 设计研制了1台结构简单、操作方便、样品池内径185 mm和高200 mm用于钚热功率测量的量热计。性能测试表明:钚量热计的样品热功率测量下限为4 mW、测量上限高于9 W,灵敏度约为202 mV·W^(-1),线性范围为4 mW~9 W。使用该量热计测量了钚样品的热功率,并与用分析化学和称量法测得的样品热功率进行了比较,结果表明:该量热计测量钚样品热功率具有较高的测量精度,当样品热功率不小于200 mW时,精密度优于0.5%。 A simple-structure plutonium calorimeter was developed and the size of sample cell is φ185 mm and high 200 mm. The performance tests show that the sensitivity of calorimeter is about 202 mV · W-1 and the linear scope of the measuring thermal power is from 4 mW to 9 W. The thermal power of plutonium was measured by the calorimeter. And it was compared with that from chemical analysis and weighting method. The results show that a good precision could be expected to measure thermal power of plutonium sample by this apparatus. The precision of the calorimeter is better than 0. 5%, when the thermal power of sample is not lower than 200 mW.
出处 《原子能科学技术》 EI CAS CSCD 2003年第z1期161-165,172,共6页 Atomic Energy Science and Technology
关键词 量热计 比对 calorimeter plutonium comparison
  • 相关文献

参考文献5

  • 1[1]Kapulla H,Heine R.Experimental Results and Experience With the KfK-Tritium Calorimeter[J]. Fusion Technol,1992,28:1 020~1 025.
  • 2[2]Mason JA,Vassallo G.Tritium Measurement by Isothermal Calorimetry[J].Fusion Technol,1992,21:425~429.
  • 3[3]Thornton MI,Vassallo G,Miller J, et al.Design and Performance Testing of a Tritium Calori-merter [J]. Nucl Instrum Methods Phys Res Sect A,1995,363:598~603.
  • 4[4]Hans A,Reinhard K,Rolf H.Tritium Inventory Measurement Using Calorimetry[J]. Fusion Technol,1992,21:412~418.
  • 5[5]ANSI N15.22-1987,Plutonium-bearing Solids Calibration Techniques for Calorimetric Assay[S].

同被引文献23

  • 1陈明秀,赵志平.YTH型液体闪烁氚测量仪的设计[J].核电子学与探测技术,2004,24(6):721-725. 被引量:6
  • 2师学明,刘成安.符合计数在钚的属性测量中的应用研究[J].中国工程物理研究院科技年报,2004(1):380-380. 被引量:1
  • 3郭文胜,邹乐西,李洁,马晓,孙天河.环境大气氚的测定技术研究[J].原子能科学技术,2007,41(1):125-128. 被引量:11
  • 4庹先国,穆克亮,雷家荣,李向阳,杨雪梅.高探测效率氚测量仪[J].核技术,2007,30(4):388-390. 被引量:8
  • 5R Coll'e.Classical radionuclidic calorimetry[J].Metrologia,2007,44:S118-126.
  • 6Mason JA,Vassallo G.Tritium Measurement by Isothermal Calorimetry[J].Fusion Technol,1992,21:425-429.
  • 7Ramthun H.Recent developments in calorimetric measurements of radioactivity[J].Nuclear Instruments and Methods,1973,112:265 -272.
  • 8Manes K R, Simmons W W. Statistical optics applied to high-power glass lasers[J]. J Opt Soc Am A, 1985, 2(4): 528-538.
  • 9Laabs H, Gao C, Weber H. Twisting of three-dimensional Hermite-Gaussian beams[J]. J Mod Opt, 1999, 46(4): 709-719.
  • 10Simon R, Sudarshan E C G, Mukunda N. Anisotropic Gaussian Schell-model beams: passage through optical systems and associated invariants[J]. Phys Rev A, 1985, 46(4): 2419-2434.

引证文献6

二级引证文献7

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部