BC-5500五分类血细胞分析仪性能评价 被引量:3

The performance evaluation of BC-5500
下载PDF
导出
摘要 目的 对BC-5500全自动五分类血细胞分析仪的主要性能进行评价,并将其与XE-2100进行对比. 方法 依照ICSH、CLSI等标准对BC-5500的携带污染率、精密度、重复性、线性范围、可比性等项目进行评价. 结果 各参数携带污染率均小于1,所有的CBC参数具有较好的精密度与重复性,WBC、RBC、HGB、MCV、PLT的CV值均小于4%;WBC、RBC、HGB、PLT具有良好的线性(相关系数r>0.9998);与XE-2100进行对比,WBC、RBC、HGB、PLT、MCV的相关系数R均大于0.9800;白细胞分类与XE-2100比较,嗜中性粒细胞(Neu%)、淋巴细胞(Lym%)、单核细胞(Mon%)、嗜酸粒细胞(Eos%)、嗜碱性粒细胞(Bas%)的相关系数分别为0.9902、0.9901、0.9145、0.8907、0.5764. 结论 BC-5500具有良好的性能,与XE-2100具有较好的可比性. Objective To evaluate the BC-5500 auto hematology analyzer and compare its performance with the System XE-2100. Methods The BC-5500 was evaluated according to guidelines published by the International Committee for Standardization in Hematology (ICSH), Clinical and Laboratory Standards Institute (CLSI). Results The results demonstrated the carryover of all CBC parameters were less than 1%; all CBC parameters had fine precision and repeatability, the CVs of WBC, RBC, HGB, MCV and PLT were less than 4%; WBC, RBC, HGB and PLT had good linearity (r>0.9998). Correlation between the BC-5500 and XE-2100 was excellent (r>0.9800) for all major CBC parameters (WBC, RBC, HGB, PLT and MCV). Correlation of the WBC differential parameters, Neu%, Lym%, Mon%, Eos% and Bas%, between the BC-5500 and XE-2100 were respectively 0.9902, 0.9901, 0.9145, 0.8907 and 0.5746. Conclusion We conclude that the overall performance of BC-5500 is excellent and closely correlates to XE-2100.
出处 《现代医学仪器与应用》 2008年第3期26-30,共5页
关键词 全自动血细胞分析仪 BC-5500 血细胞计数 分类 Auto hematology analyzer BC-5500 Blood cell count Differential
  • 相关文献

参考文献2

  • 1[1]ICSH.Clin Lab Heamotal[S].ICSH,1984,(6):69.
  • 2[2]Clinical and Laboratory Standards Institute (CLSI).Method Comparison and Bias Estimation Using Patient Samples,Approved Guideline-Second Edition.Document EP9-A2.940 West Valley Road,Suite 1400,Wayne,PA 19087-1898 USA,2004.

同被引文献24

  • 1闫虹.流式细胞术的临床应用[J].医学检验与临床,2008,20(5):4-6. 被引量:11
  • 2吴华军,王宇军,陈灵芝.UF-1000尿沉渣流式细胞仪在尿红细胞分析中的应用[J].浙江医学,2009(12). 被引量:5
  • 3黄江华 杨建琼.BC-5500血细胞分析仪全血与预稀释检测结果分析.医学与护理,2009,:58-58.
  • 4ICSH. Guidelines for the evaluation of blood cell analysers inclu ding those used for differential leucocyte and reticulocyte counting and cellmarker applications[J]. Clin Lab Haemat, 1994, 16 (2): 157-174.
  • 5National Committee for Clinical 1.aboratory Standards. Method comparison and bias estimation using patient sample[S]. Aproved Guideline EP9-A. Pennsylvania: NCCLS, 1986 : 1-14.
  • 6中华人民共和国.中华人民共和国医药行业标准:血液分析仪[M].北京:中国标准出版社,2008:61-66.
  • 7Tung J W,Heydari K,Tirouvanziam R,et al. Modem flow cytometry: a practical approach[J]. Clin Lab Med,2007,27(3) :453 - 468.
  • 8D 'Urso V, Collodoro A, Mattioli E, et al. Cytometry and DNA ploidy : clinical uses and molecular perspective in gastric and lung cancer[J]. J Cell Physiol,2010,222(3) :532 - 539.
  • 9Alvarez Barrientos A, Arroyo J, Canton R, et al. Applications of flow cytometry to clinical microbiology [ J ]. Clin Microbiol Rev, 2000,13(2) :167 - 195.
  • 10Steen H B. Flow cytometry of bacteria: glimpses from the past with a view to the future[J]. J Mierobiol Methods,2000,42( 1 ) :65 - 74.

引证文献3

二级引证文献10

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部