期刊文献+

磁共振成像交替阻抗微带线射频线圈仿真分析

Simulation of alternating impedance microstrip RF coils for MRI
下载PDF
导出
摘要 为研究改变交替阻抗微带线射频线圈的几何尺寸对感兴趣区内磁场分布的影响,保持微带线射频线圈总长度不变,改变宽带与窄带的长度比例和宽度比例,在HFSS中建立交替阻抗微带线射频线圈的仿真模型,与传统微带线射频线圈进行比较,利用ADS与HFSS的协同仿真实现线圈的调谐和匹配.仿真结果表明交替阻抗微带线射频线圈感兴趣区内的磁场均值比传统微带线射频线圈的提高了一倍以上.调整交替阻抗微带线射频线圈的宽带与窄带的长度比例和宽度比例,可以提高感兴趣区内的磁场强度,同时改善磁场分布的均匀性. This paper studied the impact of modifying the geometric parameters of alternating impedance microstrip radio frequency(RF) coils on magnetic field distribution in regions of interest.Keeping the total length of microstrip RF coils unchanged,we altered the ratio of the length and the width of both the thick and the thin sections.Simulation models in HFSS were then built and compared with conventional microstrip RF coils.Coils were tuned and matched to 50 Ω coaxial cable with ADS and HFSS co-simulation.The results showed that the mean magnetic field intensity of alternating impedance microstrip RF coils in regions of interest was twice above that of conventional microstrip coils.Fine adjustment of the length ratio and the width ratio of the thick and the thin sections of alternating impedance microstrip RF coils can increase magnetic field intensity and improve the uniformity of magnetic field distributions.
出处 《中国计量学院学报》 2013年第1期40-43,76,共5页 Journal of China Jiliang University
基金 国家自然科学基金资助项目(No.81101031 30900332) 国家质检公益性行业科研资助项目(No.201210097) 浙江省科技厅项目(No.2010C14010) 浙江省新苗人才计划项目(No.2012R409043)
关键词 磁共振成像 交替阻抗 微带线 射频线圈 magnetic resonance imaging(MRI) alternating impedance microstrip RF coil
  • 相关文献

参考文献11

  • 1张宏杰,宋枭禹,包尚联,俎栋林.磁共振成像射频线圈技术[J].中国医学影像技术,2005,21(9):1440-1441. 被引量:24
  • 2Jin J M. Electromagnetic Analysis and Design in Magnetie Resonance Imaging[M].New York:crc Press,1999.137-138.
  • 3Nasimuddin. Microstrip antennas[M].Janeza Trdine:InTech,2011.1-2.
  • 4JamesJR,HallPS,WoodC. Microstrip Antenna Theory and Design[M].London:Peter Peregrinus Ltd,1981.12-18.
  • 5Vaughan J T,Adriany G,Strup J. Parallel transceiver for nuclear magnetic resonance system[P].US,6,969,992,2005.
  • 6Adriany G,Van De M P F,Wiesinger F. Transmit and receive transmission line arrays for 7 tesla parallel imaging[J].Magnetic Resonance in Medicine,2005.434-445.
  • 7Akgun C E,Delabarre L,Snyder C J. Novel multichannel transmission line coil for high field magnetic resonance imaging[A].Boston,MA,United States:IEEE,2009.1425-1428.
  • 8Akgun C E,Delabarre L,Snyder C J. Alternating impedance multi-channel transmission line resonators for high field magnetic resonance imaging[A].Anaheim,CA,United States:IEEE,2010.756-759.
  • 9Elabyad I A,Omar A. An investigation of alternating impedance microstrip transceiver coil arrays for MRI at 7T[A].Baltimore,MD,United States:IEEE,2011.5972857.
  • 10宋旭亮,朱义胜.微带天线的设计和阻抗匹配[J].现代电子技术,2008,31(1):73-75. 被引量:24

二级参考文献15

  • 1林昌禄.近代天线设计[M].北京:人民邮电出版社,1987..
  • 2Doty FD, Entzminger G, Yang YA. Magnetism in HR NMR probe design[J]. Concepts Magn Reson, 1998,10(4) :133 156.
  • 3Harald HQ, Mark EL, Gesine G, et al. Single-loop coil concepts for intravascular magnetic resonance imaging [J] . MRM, 1999,41 (4) :751-758.
  • 4Tropp J. The theory of the bird-cage resonator[C].In Proceedings of the 11th Annual Meeting of the Society of Magnetic Resonance in Medicine. Society for Magnetic Resonance in Medicine, 1992.
  • 5Su S, Zou MX, Murphy-Boesch J. Solenoidal array coils [J]. MRM, 2002,47(4): 794-799.
  • 6Jin JM. Electromagnetics in magnetic resonance imaging[J]. IEEE Antennas and Propagation Magazine, 1998,40( 1 ): 7-22.
  • 7Doty FD. Probe design and construction [J]. Encyclopedia of NMR, 1996,6(2): 3753-3762.
  • 8Wright SM, Wald LL. Theory and application of array coils in MR spectroscopy[J]. NMR Biomed, 1997,10(6) :394-410.
  • 9Vaughan JT, Hetherington HP. High frequency volume coils for clinical NMR imaging and spectroscopy [J] . MRM, 1994, 32 ( 2 ): 206-218.
  • 10Pearce RB. New surface coil designs enhance MRI performance[J]. Diagn Imaging, 1988,10(1): 117-129.

共引文献46