期刊文献+

PIN探测器对脉冲辐射响应的电荷线性特性研究

Investigation on Linear Charge Output of PIN Detectors for Pulsed Radiation
下载PDF
导出
摘要 采用理论模拟和IVA(inductive voltage adder)产生的脉冲X射线,研究了PIN探测器输出电荷随脉冲辐射强度的变化关系。结果表明,当入射脉冲辐射注量是PIN探测器最大线性电流对应注量的16.2倍时,PIN探测器输出电荷仍然与脉冲辐射强度呈线性变化,偏离线性小于5%。 The relationship between the output charge of PIN detectors and the intensity of the pulsed X-rays was investigated numerically and experimentally on an inductive voltage adder(IVA).It is shown that the output charge of PIN detectors remains linear when the fluence of pulsed X-rays is 16.2times as large as that causing current saturation in the detec- tors.The deviation of the charge linearity is less than 5%.
出处 《现代应用物理》 2013年第3期232-235,288,共5页 Modern Applied Physics
关键词 PIN探测器 线性电荷 脉冲辐射测量 PIN detector linear charge pulsed radiation measurement
  • 相关文献

参考文献9

  • 1KNOLL G F. Radiation Detection and Measurement[M].New York:John Wiley & Sons,2000.
  • 2MESQUITA C H,FILHO T M,HAMADA M M. Development of neutron detector using the pin photodiode with polyethylene (n,p) converter[J].{H}IEEE Transactions on Nuclear Science,2003,(04):1170-1174.
  • 3刘庆兆.脉冲辐射场诊断技术[M]{H}北京:科学出版社,1994.
  • 4欧阳晓平,李真富,王群书,霍裕昆,马彦良,张前美,张国光,金玉仁.狭缝式高灵敏裂变中子探测系统[J].物理学报,2005,54(10):4643-4647. 被引量:17
  • 5张显鹏,欧阳晓平,张忠兵,田耕,陈彦丽,李大海,张小东.组合式Si-PIN 14 MeV中子探测器[J].物理学报,2008,57(1):82-87. 被引量:12
  • 6GLEBOV V Y,MEYERHOFER D D,SANGSTER T C. Development of nuclear diagnostics for the National Ignition Facility[J].{H}Review of Scientific Instruments,2006,(10):10E715-10E1-7.
  • 7WARD A L. Effect of space charge upon the transport of charge carriers[J].{H}Journal of Applied Physics,1964,(03):471-475.
  • 8张忠兵,张显鹏,陈亮,欧阳晓平.PIN探测器贯穿辐射与非贯穿辐射线性特性[J].原子能科学技术,2010,44(5):595-599. 被引量:3
  • 9郭洪生,何锡钧,彭太平.大面积PIN探测器性能参数实验测量[J].原子核物理评论,2005,22(3):272-275. 被引量:3