摘要
讨论了氟电极的空白电位值及其对标准曲线的影响 ,提出了用标准校正法消除空白电极电位漂移对分析结果产生影响的方法 ,使分析结果的精密度和准确度大大提高 ,为准确测定水中的F-提供了一种简便可靠的方法。
The blank value on electric potential of fluoric electrode and its influence on standard curve were discussed.The method which eliminates influence of analysis result caused by the blank electrode electric potential drift was proposed, which increases the precision and accuracy considerably. This paperpresents, a brief and dependable approach.
出处
《山东农业大学学报(自然科学版)》
CSCD
北大核心
2004年第2期251-253,共3页
Journal of Shandong Agricultural University:Natural Science Edition