摘要
该文研究了基于氮化钛的银离子选择电极的制备及其响应性能。研究结果表明该电极对Ag^+的响应斜率约为60 mV/dec,线性范围为1×10^(-2) mol/L~1×10^(-5) mol/L,检测下限为6.3×10^(-6) mol/L。
Titanium nitride was recommended as silver-sensitive materials to fabricate sensors. The electrode shows a linear response in the concentration range of 1×10^(-2) mol/L~1×10^(-5) mol/L with a slope of 60 mV/decade. The detection limit is 6. 3×10^(-6) mol/L. It exhibits good selectivity toward common ions. The electrode also shows a high stability and good reproducibility.
出处
《化学传感器》
CAS
2004年第3期23-25,共3页
Chemical Sensors