摘要 目的研究血清样本稀释度、封闭液、孵育条件、洗涤条件及酶标板包被后保存时间对血清抗蛋白酶3(PR3)抗体检测结果的影响。方法采用酶联免疫吸附试验(ELISA)对159份血清进行抗PR3抗体检测,并用受试者工作特征曲线(ROC曲线)分析确定阳性界限(cut-off)值。结果血清样本最佳稀释度为1∶50;5%脱脂奶粉封闭酶标板能有效地降低本底值;孵育缓冲液中加入牛血清白蛋白及脱脂奶粉比单用磷酸盐缓冲液(PBS)对测定值有明显影响。经ROC曲线分析实验的阳性临界值为正常阴性对照吸光度(A)值的2.5倍,灵敏度和特异性分别为84.62%和98.46%。结论血清抗PR3检测结果受血清样品稀释、封闭液、孵育条件及阳性临界值的选择等多因素的影响。 Objective To investigate the influence factors on the detective level of serum anti-proteinase 3 (anti-PR3) antibody by analyzing the absorptance at various conditions, including sample dilutions, blocking materials, incubation condition and interval time between blocking and detection. Methods Anti-PR3 from 159 serum samples (13 patients and 146 controls) was assayed by ELISA, and the cut-off values were defined by ROC curve obtained. Results Serum dilution at 1∶50 was suitable for the detection; absorptance of background blocked by 5% defatted milk powder was decreaced significantly; the incubation buffer containing bovine serum albumin or 5% defatted milk powder influenced the detection more obviously than that containing PBS only; the area under curve (AUC) was 0.928 (95% CI 0.834~1.021); cut-off value of absorptance was 0.235; sensitivity and specificity were 84.62% and 98.46% respectively. Conclusion The detection of anti-PR3 can be influenced by serum dilution, blocking materials, incubation condition and the choice of cut-off point.
出处 《中国药物与临床》 CAS 2004年第10期752-754,共3页 Chinese Remedies & Clinics
基金 山西省高校科技研究开发基金资助项目(2001-10)
关键词 血清 抗蛋白酶 阳性 PR 抗体检测 孵育 酶联免疫吸附试验(ELISA) 封闭液 包被 稀释度 Anti-proteinase 3 Enzyme-linked immunosorbent assay Cut-off value
  • 相关文献

参考文献5

  • 1[1]Krestin WA, Daina S, Per B, et al. Clinical evaluation of a capture ELISA for detection of proteinase-3 antineutrophil cytoplasmic antibody. Kidney Int, 1998,53:1230-1236.
  • 2[2]Zhao MH,Jones SJ,Lockwood CM. Bactericidal/permeability-incresing protein (BPI) is an important antigen for anti-neutrophil cytoplasmic antibodies (ANCA) in vasculitis. Clin Exp Immunol,1995,99: 49-56.
  • 3[3]Baslund B,Segelmark M,Wilk A, et al. Screening for anti-neutrophil cytoplasmic antibodies (ANCA): is indirect immunofluorescence the method of choice? Clin Exp Immunol, 1995,99:486-492.
  • 4[4]Wang G,Csernok E, de Groot K,et al. Comparison of eight commercial kits for quantitation of antineutrophil cytoplasmic anti bodies (ANCA). J Immunol Method, 1997,208(2):203-211.
  • 5[5]Hagen EC, Andrassy K, Csernok E, et al. Development and standardization of solid phase assays for the detection of anti-neutrophil cytoplasmic antibodies (ANCA): a report on the second phase of an international cooperative study on the standardization of ANCA assays. J Immunol Method, 1996, 196: 1-15.