摘要
目的 探讨血小板体外保存指标 ,验证国产血小板保存箱实际保存效果。方法 取新分离的富血小板血浆 (PRP) 12袋 ,每袋平均分成两袋 ,分别放入国产XHZ IA型血小板保存箱和进口FORMA 36 0 6型血小板保存箱中保存 5天 ,每天取样检测血小板聚集功能、低渗休克反应回复功能、CD6 2p表达率等体外保存指标。结果 两种保存箱保存PRP各项体外保存指标无显著差别。结论 XHZ IA血小板保存箱与FORMA血小板保存箱保存PRP的实际保存效果无差别。
Objective To study the preservative indices of stored platelet and verify the effect of a new platelet incubator made in China. Methods 12 bags of fresh platelet rich plasma (PRP) were collected. Each bag was equally divided into 2.One bag was stored for 5 days in XHZ-IA platelet incubator which was made in China, and the other bag stored in FORMA-3606 platelet incubator made in USA. Some preservative indices of stored platelet e.g. aggregation rate, recovery rate after response to hypotonic shock, CD62p expression rate etc. were tested daily. Results There were no significant differences in these preservative indices between platelets stored in two platelet incubatorshas. Conclusion XHZ-IA platelet incubator has the same effect on PRP preservation as FORMA-3606 platelet incubator.
出处
《中国输血杂志》
CAS
CSCD
2004年第5期301-303,共3页
Chinese Journal of Blood Transfusion
基金
"十五"军队医药卫生科研基金 ( 0 1L0 45 )资助项目