摘要 The prototype neutron flux monitor consists of a high purity ^(235)U fission chamber detector and a'blank'detector,which is a fissile material free detector with the same dimension as the fission chamber detector to identify noise issues such as noise coming from gamma rays.The main parameters of the fission chamber assembly that have been measured in the laboratory are confirmed to approach the technological level of the International Thermonuclear Experimental Reactor(ITER)in the near future.This prototype neutron flux monitor will be further developed to become a neutron flux monitor suitable for the operation phase of D-D fusion on the ITER.
作者 Yang Jinwei Song Xianying Zhang Wei Li Xu Lee Wenzhong Wang Shiqing Xiao Gongshan Yang Bo Lu Shuangtong 杨进蔚;宋先瑛;张炜;李旭;郦文忠;王世庆;肖功珊;杨波;陆双桐(Southwestern Institute of Physics,Chengdu 610041,China;Beijing Nuclear Instruments Factory,Beijing 100020,China)
出处 《Plasma Science and Technology》 SCIE EI CAS CSCD 2005年第3期2860-2862,共3页 等离子体科学和技术(英文版)
基金 The project supported by the National Natural Science Foundation of China(No.10175021) the Retraining Foundation of the Southwestern Institute of Physics for Talented Personnel
  • 相关文献

参考文献5

  • 1Nishtani T, Kaasain S, Johnson L G, et al. Rev. Sci.Instrum., 1999, 70 (1): 1141.
  • 2Krasilnikov A V. Status of RF(Russia Federacy) Activity in Neutron Diagnostic Development for ITER,Sixth Meeting of the ITPA Topical Group on Diagnostics, Naka, 19-21 February 2004.
  • 3Nishtani T, Takeuchi H, Kondoh T, et al. Rev. Sci.Instrum.,1992, 63 (11): 5270.
  • 4Ji Changsong. Handbook of Nuclear Radiation Detectors & Their Experimental Techniques First edition,Beijing: Atomic Energy Publishing House, 1990, 13(in Chinese).
  • 5ITER Design Description Document Diagnostics,5.5.B Neutron diagnostics.

同被引文献11

  • 1ITER Council. ITER - Final Design Report [ R ]. 2001.
  • 2V. Khripunov. Nuclear performance of the D - D phase of ITER [J]. Fusion Engineering and Design, 2000,51 - 52:281 - 287.
  • 3V. Khripunov. ITER first wall as a source of photon - neutrons [ J ]. Fusion Engineering and Design, 2001, 56 - 57:899 - 903.
  • 4ITER Design Description Document Diagnostics, 5.5. B Neutron diagnostics.
  • 5A. V. Krasilnikov, M. Sasao, Yu. A. Kaschuck, et al. Status of ITER neutron diagnostic development [ J ]. Nuclear Fusion, 2005, 45(12) :1503 - 1509.
  • 6Costley, A.E. Oct. Towards Diagnostics for a fusion reactor [ J ]. Plasma Science, IEEE Transactions on, 2010, 38(10) :2934 -2943.
  • 7WU Yichan. Monte Carlo Automatic Modeling System Code, Version 4.7 Beta, FDS Team, Reactor Tech- nology Division, Institute of Plasma Physics, Chinese Academy of Sciences,2010.
  • 8K. Asai, T. lguchi, K. Watanabe, et al. Detailed design of ex - vessel neutron yield monitor for ITER [J ]. Review of Scientific instrument, 2004, 75 (10) :3537 - 3539.
  • 9K. Asai, T. Iguchi, T. Nishitani. , et al. Improved in situ calibration technique for ITER ex - vessel neutron yield monitor [ J ]. Fusion Engineering and Design, 2006, 81, (8 -14) :1497 -1502.
  • 10王磊,王侃,余纲林.基于MPI的MCNP程序的并行计算研究[J].核电子学与探测技术,2008,28(1):163-165. 被引量:7