期刊文献+

电感耦合等离子体质谱法测定纯铝中痕量杂质 被引量:4

Study on the Determination of impurity elements in Pure Aluminum by Inductively Coupled Plasma-Mass Spectrometry
下载PDF
导出
摘要 建立了电感耦合等离子体质谱法(ICP -MS)测定纯铝中痕量铜、铁、锰、锌、钛、钴、铅、镍和铬等杂质元素的测定方法。考察了基体效应及接口效应对仪器测定灵敏度的影响,选择适当的待测元素同位素克服了质谱干扰,定量加入内标元素有效校正了基体效应、接口效应及仪器波动等因素而带来的测量偏差。结果表明,方法的检出限为0 .0 11~0 .12 μg/L ,回收率为90 .0 %~10 6.0 % ,相对标准偏差(RSD)小于3 .5 7%。该法准确、快速、简便,应用于纯铝中杂质元素的测定,结果满意。 The method of inductively coupled plasma-mass spectrometry(ICP-MS) for the determination of Cu,Fe,Mn,Zn,Ti,Co,Pb,Ni and Cr in pure aluminum was studied.The instrumental parameters and the effects of matrix and interface have been studied.The influence of mass spectrometry interference has been corrected by selecting different isotopes.The adding internal standard elements has corrected the effects of matrix and interface efficiently.The detection limit was 0.011~0.12μg/L,the recovery ratio was 90.0%~106.0%,and the RSD was less than 3.57%. The method was accurate,quick and convenient,and acceptable for determination of impurity elements with satisfactory result.
出处 《轻金属》 CSCD 北大核心 2005年第5期30-33,共4页 Light Metals
关键词 纯铝 杂质元素 电感耦合等离子体质谱 基体效应 内标元素 pure aluminum impurity elements ICP-MS effects of matrix internal standard elements
  • 相关文献

参考文献6

二级参考文献18

  • 1R K Winge 钱国贤(译).电感耦合等离子体发射光谱图册[M].北京:中国光学学会光谱学会,1986..
  • 2刘宏远,分析化学译刊,1987年,1/2卷,4期,77页
  • 3Dang H J,Z Anal Chem,1984年,318卷,1期,22页
  • 4陈新坤,电感耦合等离子体光谱法原理和应用,1987年
  • 5钱国贤(译),电感耦合等离子体发射光谱图册,1986年
  • 6汪尔康.21世纪的分析化学[M].北京:科学出版社,1998..
  • 7R Juvonen,T Lakomaa.Talanta[J] ,2002,58(3):595.
  • 8M El Himri,A Pastor,M De La Guardia.Anal.Bioanal.Chem.[J] ,2000,367(2):151.
  • 9USA EPA method 200.8 .Determination of trace elements in waters and water by inductively coupled plasma-mass spectrometry [S].1994.
  • 10GB9644—88硬聚氯乙烯(PVC—U)饮水管材和管件铅、锡、镉、汞的萃取方法及允许值[S].