摘要
本工作是在自建装置上,利用时间分辨激光荧光光谱技术,选择Eu-TTA-EtOH液体荧光体系,建立了时间分辨激光荧光测定铕的方法。该法测定铕离子的线性范围为0.0002~10μg/ml,检测限为10-5μg/ml。0.03μg/mlEu3+标准样品EuTTA3,测定的相对标准偏差为4.7%(n=12),该法用于实际样品的分析测定,结果满意。结合铕的分析,对本方法选择化学体系的条件、测定装置的参数进行了讨论。并获得了一些基本实验数据。
A variable gated photomultiplier technique capable of separating weak prompt fluorescence from the total emission of complexes is described. This technique includes the use of a N2 laser excitation source and a Boxcar integrator for sampling and time averaging the emission signals. The Boxcar enables one to pick the segments of emission at variable time delays. The EuTTA3 in EtcH emits more strong fluorescences under irradiation of N2 laser.The optimal condition for determination of Eu3+ on the set-up was studied. The linear range and detection limit of the method were 2×10-4 -10μg/ml and 1× 10-5μg/ml, respectively. The relative standard deviation for the determination of EuTTA3 (Eu3+: 0. 03μg/ml, n =12) was 4. 7%. The method is simple and highly sensitive and can be used to determine real samples.
出处
《光谱学与光谱分析》
SCIE
EI
CAS
CSCD
北大核心
1995年第4期17-21,共5页
Spectroscopy and Spectral Analysis
基金
中科院基础局和长春分院资助
关键词
激光荧光
铕
荧光光谱
Laser fluorescence, Time-resolved spectra, Europium