摘要
本文研究了铜─钙黄绿素流动注射荧光熄灭法测定铜的条件,此法操作简便、快速、重现性好,进样频率120次/h,相对标准偏差小于1.5%,检测限0.023μg/ml.用于测定牛肝、山药及污水中痕量铜,结果满意。
The method is studied for the determination of trace copper using the system of copper-calcein by fIow injection fluorescence quenching.The method proposed has the advantages of conve-nience, rapidity and reliability.The relative standard deviation is less than 1. 3% and limit of detectionis 0.023μg/ml. This method has been used for analysing liver,chinese yam and waste water.
出处
《冶金分析》
CAS
CSCD
北大核心
1995年第1期7-9,共3页
Metallurgical Analysis
关键词
铜
钙黄绿素
流动注射
荧光熄灭法
Copper.Calcein,Flow injec tion, Fluorescence quenching method