期刊文献+

三维肿瘤运动和体表标记物运动的研究 被引量:5

Research on Motion of 3D Tumor and Skin Marker
下载PDF
导出
摘要 目的:研究呼吸导致的肺部肿瘤运动和体表标记物运动轨迹之间的关系。材料与方法:我们利用模拟定位机对肺癌病人呼吸运动中肿瘤三维运动和体表标记物的运动进行了监测并处理,再对这些数据进行频域分析和秩相关分析。结果:我们得到了患者肺部肿瘤运动与体表标记物运动的波形,分析出了呼吸的频率和运动之间的相关性。结论:不同患者间随呼吸的肿瘤运动差异较大,且肺部肿瘤运动与体表标记物运动不都具有相关性. Objective: To study the relationship between lung tumor motion and skin marker motion resulted from respiration, Methods: Monitor and deal with the data of 3D tumor motion and skin marker motion during respiration of the lung cancer patient by simulator system, frequency domain analysis and rank correlation analysis are carried on those data. Results: we obtained waveform of 3D tumor motion and skin marker motion. We analysed the frequency of the respiration and the correlation between the motions. Conclusions: Great difference in tumor motion due to respiration was found among different cancer patients. And not all correlation exists between lung tumor motion and skin marker motion.
出处 《中国医学物理学杂志》 CSCD 2006年第1期1-4,共4页 Chinese Journal of Medical Physics
基金 江苏省卫生厅重点科研课题(H200331)
关键词 肺癌 肿瘤运动 模拟机 快速傅立叶变换 秩相关 Lung cancer tumor motion simulator system FFT rank correlation
  • 相关文献

参考文献7

  • 1Balter JM, Ten Haken RK, Lawrence TS, et al. Uncertainties in CT_based radiation therapy treatment planning associated with patient breathing[J].Int J Radiat Oncol Biol Phys. 1996,36:167-174.
  • 2International Commission on Radiation Units and Measuremets(ICRU)[M]. Reports 62.ICRU Publication. 1999:1097-1101.
  • 3胡广书.数字信号处理[M].北京:清华大学出版社,1997..
  • 4Jeremy D.P.Hosak, M.SC., Katharinae.sixel, PH.D. Correlation of lung tumor motion with external surrogate indicators of resiration[J]. Radiation Oncology Biol. Phys. 2004,60(4):1298-1306.
  • 5Ekberg L, Holmberg O, Witgren L, et al. What margins should be added to the clinical larget volume in radiotherapy treatment planning for lung cancer[J]. Radiother Oncol.1998,48:71-77.
  • 6Yvete S, Hiroki S, et al. Precise and real-time measurement of 3D tumor in lung due to breathing and heartbeat, measured during radiotherapy [J]. Int J Radiat Oncol Biol Phys 2002,53(4):822-834.
  • 7Yoshikazu Tsunashima, M.SC,, Takeji Sakae, PH.D,, Correlation between the respiralory waveform measured using a respiratory sensor and 3D tumor motion in gated radiotherapy [J]. Radiation Oncology Biol. Phys,2004,60(3):951-958.