摘要 使用自行设计的超高真空金属系统和气体样品净化装置,以引进的Gastrace-s型气体分析四极质谱计为分析器,以纯度为999%的3He和4He气体为标准气样,对微量气体样品中He同位素的四极质谱峰高比测定法进行了实验研究。结果表明,Zr-Al活性气体吸收剂能在390s内将活性杂质气体含量为40.7%的3He气体样品净化到接近100%。当He进样量>5×10-7ml.atm时,仪器灵敏度稳定,S(3He)/S(4He)=0.41,在本系统中,仪器对He原子的探测极限达到了1010。 Experiments are made for measurement 3He release from metal tritides by using ultravacuum metal system and gas purification line which is designed by the authors. 3He analysis are performed by a set of Gastrace quadrupole mass spectrometer (QMS). With 3He and 4He standards of 99. 9 %purity, the QMS peak height ratio of helium isotopes in trace gaseous samples are measured. Results show that Zr-Al gas getter can purificate out any impurities in 3He sample, and the purity of sHe can be increased to nearly 100 % from 40. 7 % within times of 390 s. When the input quantity of He is more than 5 × 10-7 ml' atm, the sensitivity of QMS can be maintained stably, and S3He/S (4He)equal to 0. 41. The detec limitation of such system for He is 1010 atoms.
出处 《原子能科学技术》 EI CAS CSCD 北大核心 1996年第1期55-61,共7页 Atomic Energy Science and Technology
基金 中国工程物理研究院科学基金
关键词 氦同位素 氚化物 四极质谱法 Helium isotope Metal tritides Quadrupole mass spectrometric