期刊文献+

基于平衡取样电路的高速脉冲发生器 被引量:6

High speed pulse generator based on the balance sampling circuit
下载PDF
导出
摘要 当宽带取样示波器的平衡取样电路直流偏置电压Voffset不为零时,在选通脉冲的作用下该平衡取样电路将产生一个幅度与直流偏置电压Voffset成正比的高速脉冲。文中详细地分析了高速脉冲与直流偏置电压Voffset、选通脉冲的关系。并针对一个具体的平衡取样电路进行了仿真研究。结果表明,平衡取样电路能够构成一个可控的高速脉冲发生器。 The balance sampling circuit is produced a high speed pulse by the action of strobe pulse when the dc biased voltage Voffset, of balance sampling circuit on the broadband sampling oscilloscope is not equal to zero, and the high speed pulse is proportion to the dc biased voltage Voffset. The relationship of high speed pulse on the dc biased voltage Voffset, and the strobe pulse is researched detailedly in this paper. Further more, through the simulation on a particular balance sampling circuit, it is shown that the balance sampling circuit is constructed a controllable high speed pulse generator.
出处 《电子测量技术》 2006年第2期7-10,共4页 Electronic Measurement Technology
关键词 平衡取样电路 高速脉冲发生器 宽带取样示波器 balance sampling circuit, high speed pulse generator, broadband sampling oscilloscope.
  • 相关文献

参考文献5

  • 1K. Rush, S. Draving, J. Kerley. Characterizing high-speed oscilloscopes [J ]. IEEE Spectrum,IEEE Press. 1990 (9): 38-39.
  • 2J. Verspeeht, K. Rush. Individual Characterization of Broadband Sampling Oscilloscopes with a " Nose-to-Nose" Calibration Procedure [J]. IEEE Transaction on Instrumentation and Measurement. 1994, 43 (2):347-354.
  • 3J. Verspecht. Broadband Sampling Oscilloscopes Characterization with the " Nose-to-Nose "Calibration Procedure. A theoretical and Practical Analysis [J]. IEEE Transaction on Instrumentation and Measurement. 1995, 44 (6):991-997.
  • 4S. M. RIAD. Modeling of the HP21430A feedthrough wide-band ( 28-ps ) sampling head [ J ].IEEE Transactions on Instrumentation and Measurement, 1982, 31 (2): 110-115.
  • 5朱江淼,刘明亮,卢峰.Nose-to-nose校准技术的理论分析[J].电子学报,2005,33(3):480-483. 被引量:10

二级参考文献10

  • 1Ken Rush,Steve Draving,John Kerely.Characterizing high-speed oscilloscopes[J].IEEE Spectrum September 1990.38-39.
  • 2Jan Verspecht.Broadband sampling oscilloscopes characterization with the nose-to-nose calibration procedure:A theoretical analysis[J].IEEE Transaction on Instrumentation and Measurement,1995,44(6):991-997.
  • 3Dylan F Williams,Kate A Remley.Analytic sampling-circuit model[J].IEEE transactions on microwave theory and techniques,2001,49(6):1013-1019.
  • 4Donald C DeGroot,Paul D Hale.Analysis of Interconnection Networks and Mismatch in the Nose-to-Nose Calibration[C].55th ARFTG Conference Digest.2000.116-121.
  • 5SEDKI M RIAD.Modeling of the HP-1430A feed-through wide-band (28-ps) sampling head[J].IEEE Transactions on Instrumentation and Measurement,1982,31(2):110-115.
  • 6Zhu Jiangmiao,Liu Mingliang,Gao Yun.The Simulation of the Nose-to-nose Calibration Procedure[C].Conference Proceedings of the Sixth International Conference on Electronic Measurement & Instruments,2003,2:1217-1220.
  • 7Donld R Larson,Nicholas G Paulter.The Effect of Offset Voltage on the Amplitude and Bandwidth of Kick-out Pulses Used in the Nose-to-Nose Sampler Impulse Characterization Method[C].in Proc.IEEE IMTC,Baltimore,MD,2000.1425-1428.
  • 8林茂六,赵华.宽带取样示波器Nose-to-Nose校正理论的新进展[J].电子学报,2001,29(3):409-411. 被引量:9
  • 9刘明亮,林茂六,贺云辉,高剑.基于“NTN”技术的宽带取样示波器的校准[J].计量学报,2002,23(1):65-68. 被引量:14
  • 10朱江淼,刘明亮,陈建新.用于宽带取样示波器的Nose-to-nose校准技术[J].应用基础与工程科学学报,2003,11(1):94-97. 被引量:6