摘要
本文应用DCP-AES法测定地质样品中微量锂。讨论了基体效应和阴离子效应、基体(Fe、Al、Ca、Mg、K和Na)对Li元素谱线产生不同程度的增强效应。在待测溶液中加入一定量的基体元素,能有效地降低或消除这种影响。基体元素产生的增强效应主要是阳离子的作用。还讨论了分析线的线性状况。方法简便可靠,Li的测定范围为1~1500μg/g,相对标准偏差为3.1~6.3%。
A DCP-AES method is proposed for the determination of the trace Li in geological specimen.The effects of matrix and negative ion are proposed.Matrix(Fe,Al,Ca,Mg,K and Na) brings enhanced effects of different extent to the spectrum of the element Li. The matrix interference can be eliminated effectively by adding adequate amount of matrix element into the analyte liquor. The enhanced effect presented in the matrix element is chiefly the effect of canon.The linear situation of analytic line of the element Li is also discussed.
出处
《光谱学与光谱分析》
SCIE
EI
CAS
CSCD
北大核心
1996年第3期79-82,共4页
Spectroscopy and Spectral Analysis