在线尿素清除率监测对血液透析充分性的评价 被引量:6

Evaluation on Adequacy of Hemodialysis by On-line Clearance Monitoring
下载PDF
导出
摘要 目的探讨在线尿素清除率监测(OCM)在评价血液透析充分性中的价值。方法36例在本中心行维持性血液透析的终末期肾衰患者,采用OCM仪检测尿素清除指数(K t/V),同时在透前、透后抽血,应用尿素动力学———室可变容积模型检测K t/V计算值,对OCM检测值和K t/V计算值进行比较并进行相关性分析;对其中15例透析器复用前后的K t/V比较,探讨OCM在评价血液透析充分性中的作用。结果透析器首次使用时,患者OCM检测值与K t/V计算值相比有显著差异(1.27±0.23vs1.53±0.30,P<0.001),相关性分析显示两者存在显著的直线相关性(P<0.001);透析器首次使用和复用时,患者的OCM检测值和K t/V计算值相比,亦具有显著差异(1.31±0.22vs1.55±0.30,P<0.001)。透析器复用的15例患者中,首次使用与复用第3次和第5次相比,OCM检测值及K t/V计算值均无显著差异(P>0.05)。透析每周2次与3次患者的单次血透OCM检测值及K t/V计算值相比无显著差异(P>0.05)。结论OCM检测值低于应用尿素动力学———室可变容积模型计算的K t/v值,二者呈显著正相关,可以作为评价尿毒症血透患者血液透析充分性的一项较可靠指标。 Objective To evaluate the adequacy of hemodialysis by online clearance monitoring(OCM) in hemodialysis patients. Methods The Kt/V values of 36 maintenance hemodialysis patients in our center who were assessed by both OCM and single-pool urea kinetic model were compared, and the Kt/V values of patients in the firstuse dialyser and the reused dialyser were also compared. Results In the patients who used a new dialyser,Kt/V values were 1.27±0.23 measured by OCM ,whereas 1.53±0.30 by single-pool urea kinetic model( r = 0. 574, P 〈 0. 001). In the patients who used new dialysers and reused dialysers, Kt/V values measured by OCM and single-pool urea kinetic model also showed significant differences( 1.31±0.22 vs 1.55±0.30,P 〈0. 001 ). In the fifteen patients who applied the first, third and fifth time-used dialyser,Kt/V measured by OCM and by model had no significant differences(P 〉 0.05). In all the patients undergoing dialysis two or three times per week, Kt/V measured by OCM and by model also showed no differences (P 〉 0.05). Conclusion During hemodialysis, the OCM option provides an accurate tool for continuous online monitoring of urea clearance.
出处 《上海交通大学学报(医学版)》 CAS CSCD 北大核心 2006年第5期532-534,共3页 Journal of Shanghai Jiao tong University:Medical Science
关键词 血液透析 充分性 尿素清除指数 在线尿素清除率监测 hemodialysis adequacy Kt/V on-line clearance monitoring
  • 相关文献

参考文献10

  • 1Lambie SH,Taal MW,Fluck RJ,et al.Analysis of factors associated with variability in haemodialysis adequacy[J].Nephrol Dial Transplant,2004,19 (2):406-412.
  • 2Daugirdas JT.Second generation logarithmic estimates of single-pool variable volume Kt/V:an analysis of error[J].J Am Soc Nephrol,1993,4 (5):1205-1213.
  • 3Gotch F,Levin NW,Port FK,et al.Clinical outcome relative to the dose of dialysis is not what you think:the fallacy of the mean[J].Am J Kidney Dis,1997,30(1):1 -15.
  • 4Bankhead MM,Toto RD,Star RA.Accuracy of urea removal estimated by kinetic models[J].Kidney Int,1995,48(3):785 -793.
  • 5Depner TA,Greene T,Gotch FA,et al.Imprecision of the hemodialysis dose when measured directly from urea removal.Hemodialysis Study Group[J].Kidney Int,1999,55(2):635 -647.
  • 6Lai YH,Guh JY,Chen HC,et al.Effects of different sampling methods for measurement of post dialysis blood urea nitrogen on urea kinetic modeling derived parameters in patients undergoing long-term hemodialysis[J].ASAIO J,1995,41 (2):211 -215.
  • 7Daugirdas JT,Burke MS,Balter P,et al.Screening for extreme postdialysis urea rebound using the Smye method:patients with access recirculation identified when a slow flow method is not used to draw the postdialysis blood[J].Am J Kidney Dis,1996,28(5):727 -731.
  • 8Kuhlmann U,Goldau R,Samadi N,et al.Accuracy and safety of online clearance monitoring based on conductivity variation[J].Nephrol Dial Transplant,2001,16(5):1053 -1058.
  • 9Sherman RA,Cody RP,Rogers ME,et al.The effect of dialyzer reuse on dialysis delivery[J].Am J Kidney Dis,1994,24(6):924-926.
  • 10Kopple JD.National kidney foundation K/DOQI clinical practice guidelines for nutrition in chronic renal failure[J].Am J Kidney Dis,2001,37(1 Suppl 2):S66-S70.

同被引文献46

  • 1谢华,林洪丽,李平,陈丽光,王松岩,冯青.联机尿素清除率监测和常规采血计算Kt/V对血液透析充分性评估的对照研究[J].中国血液净化,2004,3(11):587-589. 被引量:22
  • 2杨莉,王梅,徐莉.在线清除监测在评价血透患者透析剂量中的应用[J].中国医师进修杂志(内科版),2006,29(4):40-43. 被引量:12
  • 3付荣国,王莉,郭蕊军,张红利,赵莉,郭芳丽.在线血液透析滤过对终末期肾衰竭高血压患者的治疗作用[J].临床内科杂志,2006,23(5):322-324. 被引量:24
  • 4Carl D E, Feldman G. Estimating dialysis adequacy using ionic dialysance. Ren Fail, 2008, 30. 491-498.
  • 5Daugirdas J T, Schneditz D. Overestimation of hemodialysis dose depends on dialysis efficiency by regional blood flow but not by conventional two pool urea kinetic analysis. ASAIO J, 1995, 41:M719- M724.
  • 6Prado M, Roa L M, Palma A, et al. Double target comparison of blood-side methods for measuring the hemodialysis dose. Kidney Int, 2005, 68. 2863-2876.
  • 7Ficheux A, Gayrard N, Szwarc I, et al. Use of spent dialysate analysis to estimate blood levels of uraemic solutes without blood sampling: urea. Nephrol Dial Transplant. 2010, 25: 873-879.
  • 8Lira P S, Lee H P, Kho B, et al. Evaluation of pre and postdilutional on-line hemodiafihration adequacy by partial dialysate quantification and on-line urea monitor. Blood Purif, 1999, 17: 199-205.
  • 9Daugirdas J T. Second generation logarithmic estimates of single pool variable volume Kt/V: an analysis of error. J Am Soc Nephrol, 1993, 4:1205- 1213.
  • 10Daugirdas J T, Depner T A, Gotch F A, et al. Comparison of methods to predict equilibrated Kt/V in the HEMO Pilot Study. Kidney Int, 1997, 52:1395- 1405.

引证文献6

二级引证文献20

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部