期刊文献+

液体推进剂偏二甲肼自动电位滴定分析方法的建立 被引量:4

The Establishment of the Automatic Potentiometry for Unsymmetrical Dimethylhydrazine Analysis
下载PDF
导出
摘要 设计了M etrohm 809型自动电位滴定仪分析液体推进剂偏二甲肼的应用程序,解决了程序设计中滴定模式和终点确定方式选择、滴定速度确定等关键性技术问题,考察了环境温度对分析结果的影响,确定了最佳条件,实现了液体推进剂偏二甲肼的自动电位滴定分析。重复性试验和准确度试验表明,自动电位滴定法分析偏二甲肼具有准确度高、精密度好等优点,测量结果完全符合G JB 753-89的要求。 A practical program was designed for a model Metrohm 809 automatic potentiometric titratior to analyze the liquid propellant unsymmetrical dimethylhydrazine(UDMH). The key technique problems, such as titration mode, end-point identification and titration rate, were successfully resolved. The influence of environmental temperature on the experimental values was investigated and the optimum conditions were defined. Thus, the automatic potentiometry for UDMH analysis was achieved. The good experimental results with excellent precision and accuracy indicate that the proposed method is quite suitable for UDMH analysis, which is in good agreement with those by GJB 753 -89 standard method.
出处 《分析测试学报》 CAS CSCD 北大核心 2006年第4期111-114,共4页 Journal of Instrumental Analysis
关键词 液体推进剂 偏二甲肼 电位滴定 程序设计 Liquid propellant Unsymmetrical dimethylhydrazine Potentiometry titration Programme designation
  • 相关文献

参考文献6

二级参考文献13

  • 1佚名.航天发射场化验员手册[M].国防科工委后勤部,1995..
  • 2佚名.液体推进剂[M].国防科工委后勤部,1979..
  • 3杜廷发.现代仪器分析[M].长沙:国防科大出版社,1991..
  • 4国防科工委后勤部.火箭推进剂监测防护与污染治理[M].长沙:国防科技大学出版社,1993.55-56.
  • 5国防科工委后勤部.火箭推进剂监测防护与污染治理[M].长沙:国防科技大学出版社,1993.61.
  • 6刘祥萱 王煊军 郭和军.气相色谱/质谱法分析液体推进剂偏二甲肼氧化产物[J].分析实验室,2003,22:481-481.
  • 7佚名,航天发射场化验员手册,1995年
  • 8杜廷发,现代仪器分析,1991年
  • 9佚 名,液体推进剂,1979年
  • 10高思秘.液体推进剂分析[M].北京:宇航出版社,1991..