期刊文献+

髋关节假体专家系统中力学评价方法的实现 被引量:1

The Implementation of Rapid Evaluation Expert System of Hip-Joint Prosthesis Selecting
下载PDF
导出
摘要 髋关节重建术中假体的选择与手术效果直接相关,为更好地利用专家的经验,提高选择假体的准确性,开发了髋关节假体优选专家系统,并引入生物力学的快速评价方法作为选择假体的力学准则.基于患者X光片的数据信息,用Matlab和VB语言进行专家库系统的整体模块开发,对力学评价模块采用Matlab与ANSYS中的APDL语言开发.经过力学仿真及分析,得到了几何相似性和力学相似性较高的有限元模型,分析结果与文献结果较为吻合,并据此提出了在专家库系统中选择假体的评价准则. Selecting prosthesis is an important job in hip replacement and in the past, surgeons did this by matching the patient's standard hip X-rays with the prosthesis model template, relying dominantly on the experience and relevant acknowledge. This type of opting has a great deal of randomness and inaccurate- ness. In order to improve the accuracy, this paper developed a rapid-evaluation expert system based on VB, Matlab language and APDL language in ANSYS software, with which the parameterized finite ele- ment model of femur and the corresponding prosthesis were realized and the stress analysis was acquired. Based on the results of the FEA, the principles of selecting prosthesis with this expert system were put forward.
出处 《上海交通大学学报》 EI CAS CSCD 北大核心 2006年第8期1452-1456,共5页 Journal of Shanghai Jiaotong University
基金 国家自然科学基金资助项目(30530230)
关键词 髋关节 假体 生物力学 快速评价 专家系统 hip joint prosthesis bio mechanics rapid-evaluation expert system
  • 相关文献

参考文献6

二级参考文献21

  • 1李永丰,王以进,杜伟明.髋关节置换前后股骨上应力的变化[J].医用生物力学,1996,11(3):140-141. 被引量:5
  • 2钱齐荣.[D].第二军医大学,1997,4:34~141.
  • 3Van Rietbergen B, Huiskes R. Load transfer and stress shielding of the hydroxyapatite-ABG hip: a study of stem length and proximal fixation [J]. J Arthroplasty ,2001 ,16:55-63.
  • 4Ramaniraka NA, Rakotomanana LR, Rubin PJ, et al Noncemented total hip arthroplasty: influence of extramedullary parameters on initial implant stability and on bone-implant interface stresses [J]. Rev Chit Orthop Reparatrice Appar Mot,2000 ,86(6) :590.
  • 5Huiskes R, Verdonschot N, Nivbrant B. Migration, stem shape, and surface finish in cemented total hip arthroplasty[J]. Clin Orthop, 1998 ,(355):103-112.
  • 6Chang PB, Mann KA, Bartel DL. Cemented femoral stem performance. Effects of proximal bonding, geometry, and neck length [J]. Clin Orthop, 1998 , (355):57-69.
  • 7JanssonV, Refior HJ Mechanical failure of the femoral component in cemented total hip replacement-a finite element evaluation [J]. Arch Orthop Trauma Surg, 1993,113(1):23-27.
  • 8Prendergast PJ, Taylor D. Stress analysis of the proximo-medial femur after total hip replacement[J]. J Biomed Eng,1990,12(5)379-382.
  • 9Halley DK, Glassman AH. Twenty- to twenty-six-year radiographic review in patients 50 years of age or younger with cemented Charnley low-friction arthroplasty. J Arthroplasty, 2003 ,18(7 Suppl 1):79-85.
  • 10Makarand GJ, Suresh GA, Freeman M, et al. Analysis of a femoral hip prosthesis designed to reduce stress shielding. J Biomech, 2000, 33: 1655-1662.

共引文献37

同被引文献3

引证文献1

二级引证文献1

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部