期刊文献+

主动脉瓣膜泵的回流

Back-flow of aortic valvo-pumps
下载PDF
导出
摘要 为了研究主动脉瓣膜泵的回流,将血泵内的叶轮换成圆柱体,使血泵失去泵血的作用,从而进行回流测试.测试中,血泵分别以15000r/min,17500r/min和20000r/min的转速旋转,出口与进口之间的压力差由一个离心泵提供,用流量计测出从瓣膜泵出口经定、转子间气隙到进口的回流.试验结果表明气隙和压差对回流有明显影响:增大气隙或增大压差都会导致回流的增加.当气隙为0.20mm,进出口压差为13.3kPa时,瓣膜泵的回流为0.6—0.8L/min.回流随时间变化的曲线图表明,在自然心脏舒张期回流产生一个波峰,而在收缩期产生一个波谷.瓣膜泵的转速对回流无明显影响,当转速增大时回流只有略微的下降. To investigate the regurgitation of aortic valvo-pump, the impeller of the pump was replaced by a cylinder and hereby the valvo-pump had lost its pumping function. The pressure head crossing the aortic valvopump rotating at 15 000 r/min, 17 500 r/min and 20 000 r/min respectively was established by a centrifugal pump. The back flow from outlet to inlet of valvo-pump via the air-gap between rotor and stator was measured. Results demonstrated that the air-gap and pressure head have remarkable effect on back-flow ; a larger air-gap or a larger pressure head will lead to a larger back-flow. With 0.20 mm airgap and 13.3 kPa pressure head, the valvo-pump has a 0.6 -0.8 L/min mean back-flow. Instantaneous back-flow measurement indicated that the back-flow value has a wave crest during diastole and a trough during systole of the natural heart. The rotating speed of the valvo-pump has no significnt effects on its back-flow. As the rotating speed increases the back-flow will decrease slightly.
出处 《江苏大学学报(自然科学版)》 EI CAS 北大核心 2006年第5期434-436,共3页 Journal of Jiangsu University:Natural Science Edition
基金 国家自然科学基金资助项目(39970736)
关键词 主动脉瓣膜泵 回流测量 气隙和压差 抗血栓形成 aortic valvo-pump regurgitation measurement different air-gap and pressure head anti-thrombogenicity
  • 相关文献

参考文献7

  • 1Qian KX, Wang DF, Topaz S, Ru WM, Zeng P, Yuan HY, Zwischenberg JB. Use of aortic valvo-pumps placed in valve annulus for long term left ventricular assist [J].4SAIO,2005,51 (6) :736 -738.
  • 2Qian KX, Ru WM, Zeng P, Yuan HY. World-first implantable aortic valvo-pump with sufficient hemo-dynamic property[J]. J Med Eng Techn, 2005,29 ( 6 ) : 302-304.
  • 3Qian KX, Ru WM, Zeng P, Yuan HY. World-smallest LVAD with 27 g weight , 21 mm OD and 5 L/min flow with 50 mmHg pressure increase[J]. J Med Eng Techn.(to be published)
  • 4钱坤喜,曾培,茹伟民,袁海宇.用于左心室辅助的φ25mm主动脉瓣膜泵[J].江苏大学学报(自然科学版),2005,26(4):361-363. 被引量:2
  • 5钱坤喜,王颢,王芳群,茹伟民,曾培,袁海宇,封志刚.21毫米人造心脏瓣膜泵的设计及研制[J].上海生物医学工程,2005,26(3):160-162. 被引量:2
  • 6钱坤喜,曾培,茹伟民,袁海宇.人造心脏泵及心脏瓣膜泵研究[J].中国医疗器械杂志,2005,29(4):238-240. 被引量:4
  • 7Qian KX. Novel magnetic spring and magnetic bearing[J]. IEEE Trans Magnetics, 2003,39 : 559 - 601.

二级参考文献31

  • 1钱坤喜.一种新颖的磁力弹簧[P].中国专利,No.962315532.1996.
  • 2钱坤喜.泵机合一的叶轮式心室辅助装置[P].中国专利:(982265328)..
  • 3American Heart Association. International Cardiovascular Diseases Statistics. Statistical Fact Sheet-Populations. Dallas,TX: American Heart Associations, 1-10, 2003
  • 4Goldowsky M: Mini hemoreliable axial flow LVAD with magnetic bearings. Part 1,2,3. ASAIO J. 48:96-101,2002
  • 5Mueller J, Nuesser P, Weng Y: The first ventricular cardiac assist device without any wear and tear in a clinical multi-center trial. ASAIO J.49: 162,2003
  • 6Qian KX, Zeng P, Ru WM, Yuan HY, Feng ZG, LI L: Permanent magnetic-levitation of rotating impeller: a decisive breakthrough in the centrifugal pump. J Med. Eng. Tech. 26(1): 36-38, 2002
  • 7Qian KX, Zeng P, Ru WM, Yuan HY: A novel permanent maglev impeller TAH: Most requirements on blood pumps have been satisfied. J Biomaterials Applications. 18(1): 53-62,2003
  • 8Mitamura H, Nakamura H, Okamoto E, Yozu R, Kawada S, Kim DW: Development of the Valvo Pump: An Axial Flow Pump Implanted at the Heart Valve Position. Artificial Organs. 23 (6): 566,1999
  • 9Li G, Zhao H, Zhu X, Ren B.: Preliminary in vivo study of an intra-aortic impeller pump driven by an extracorporeal whirling magnet. Artif Organs.26(10):890.2002
  • 10Yamazaki K, Umezu M, Koyanagi H, Kitamura M, Eishi K, Kawai A, Tagusari O, Niinami H, Akimoto T, Nojiri C, Tsuchiya K, Mori T, Iiyama H, Endo M: A miniature intraventricular axial flow blood pump that is introduced through the left ventricular apex. Trans ASAIO. 38:679-683,1992