摘要
A novel modified electrode for the voltammetric determination of acetaminophen has been fabricated based on electropolymerization of L-serine on a glasy carbon electrode.Voltammetric behaviour of acetminophen on the modified electrode was investigated by cyclic voltammetry and linear sweep voltammetry.It was found that the oxidation peak current of acetaminophen was enhanced greatly on the modified electrode.A linearity between the oxidation peak current and the acetaminophen concentration was obtained in the range of 1.00×10-7-2.00×10-5 mol L-1 with a detction limit of 7.00×10-8 mol L-1.The practical application of the modified electrode in the determination of acetaminophen in a paracetamol oral solution demonstrated it has good selectivity and high sensitivity.
A novel modified electrode for the vohammetric determination of acetaminophen has been fabricated based on electropolymerization of L-serine on a glasy carbon electrode. Voltammetrlc behaviour of acetminophen on the modified electrede was investigated by cyclic voltammetry and linear sweep vohammetry. It was found that the oxidation peak current of acetaminophen was enhanced greatly on the modified electrode. A linearity between the oxidation peak current and the acetaminophen concentration was obtained in the range of 1.00×10^-7-2. 00×10^-5 mol L^-1 with a detction limit of 7.00×10^-8 mol L^-1. The practical application of the modified electrode in the determination of acetaminophen in a paracetamol oral solution demonstrated it has good selectivity and high sensitivity.
出处
《化学研究与应用》
CAS
CSCD
北大核心
2006年第9期1095-1098,共4页