期刊文献+

双通道红外光谱辐射计的研制及标定 被引量:2

Development and calibration of infrared double-channel spectral radiometer
原文传递
导出
摘要 研制了一台双通道(1.40±0.02μm,4.50±0.03μm)光谱辐射亮度计(以下简称辐射计),并进行了光谱辐亮度定标[1]。该辐射计主要由前置光学系统、精密机械调制器、双路单元红外探测器、锁相放大器、A/D转换器和单片机等组成。在短波红外波段,由InGaAs探测器和积分球光源传递国家光谱辐照度标准灯的标准[15],对辐射计进行标定[7];在中波红外波段,用大面积标准黑体辐射源和腔型黑体辐射源,标定辐射计的光谱辐亮度。由数据统计分析,得出辐射计的光谱辐亮度响应度的不确定度[12,14]。 A spectral radiation instrument (which was called radiometer for short below) with double channels (1.40 ± 0. 02μm, 4.50± 0.03μm) was developed, and its spectral radiance was calibrated. This radiometer mainly consisted of optical system, fine mechanical modulator, dual channel infrared detectors, lock-in amplifier, A/D conversion device and monolithic processor, etc. The radiometer was respectively calibrated in short-infrared band by national standard spectral irradiance lamp and in mid-infrared band by large area standard black body. Based on the statistical analysis of the data, the indeterminacy of the spectral radiance response of the radiometer was calculated.
出处 《光学技术》 EI CAS CSCD 北大核心 2006年第5期773-774,778,共3页 Optical Technique
关键词 红外辐射计 辐射定标 光谱辐亮度 响应度 infrared spectral radiometer radiation calibration radiance response
  • 相关文献

参考文献12

  • 1格鲁姆 F,贝彻雷 R J.辐射度学[M].缪家鼎,等译.北京:机械工业出版社,1987.
  • 2王之江.光学技术手册[M].北京:机械工业出版社,1987..
  • 3Kohler R,Goebel R,Pello R.Results of an international comparison of spectral responsivity of silicon photodetectors[J].Metrologia,1996,32(6):463-468.
  • 4郑小兵,吴浩宇,章骏平,刘玉成,周威,王乐意,乔延利.高精度光辐射定标和标准传递方法[J].科学通报,2000,45(12):1341-1344. 被引量:24
  • 5Foukal P V,Kochling H,Miller P.Cryogenic absolute radiometer as laboratory irradiance standards,remote sensing detector and pyroheliometers[J].Appl Opt,1990,29(7):988-993.
  • 6徐明,高岳,王霞,张化朋,高稚允,王仲春,张建勇.光谱段积分辐射计的研制[J].光学技术,2002,28(5):472-474. 被引量:1
  • 7Hoyt C C,Foukal P V.Cryogenic radiometers and their application to metrology[J].Metrologia,1991,28:163-167.
  • 8Kostkovski H J.Reliable Spectorscopy[M].Maryland:Spectroadiometry Consulting,1997.162
  • 9Gentile T R,Houston J M,Hardis J E,et al.The NIST high accuracy cryogenic radiometer[J].Appl Opt,1996,35(7):1056-1068
  • 10张建勇,金伟其,高稚允,吴玲,王霞.红外积分辐射计的测量与标定[J].北京理工大学学报,2003,23(6):749-753. 被引量:2

二级参考文献16

  • 1张敬贤 李玉丹.微光与红外成像技术[M].北京:北京理工大学出版社,1994.109-115.
  • 2何立民.MCS-51系列单片机应用系统设计[M].北京航空航天大学出版社,1985..
  • 3潘守文,现代气候学原理,1994年
  • 4王炳忠,太阳辐射能的测量与标准,1993年
  • 5白素莲,中国农业小气候研究进展,1993年,351页
  • 6国家气象局气候监测应用管理司,气象仪器和观测方法指南,1992年
  • 7William N J. Scott B. Infrared optoelectronics: Devices and applications[M]. New York: [s.n.], 1987.
  • 8Wolfe W L, Zissis G J. The infrared handbook[M]. Washington: D.C. Environmental Research Institute of Mich, 1978.
  • 9Jha A R. Infrared technology[M]. New York: Wiley, 2000.
  • 10Kostkovski H J.Reliable Spectroscopy[].Maryland: Spectroradiometry Consulting.1997