CR-39固体核径迹探测器用^(252)Cf源剂量刻度及其应用

Calibration of CR-39 solid state nuclear track detectors with a standard ^(252)Cf neutron source and its application
原文传递
导出
摘要 目的:用CR-39固体核径迹探测器测量医用电子加速器产生的污染中子辐射。方法:将CR-39固体核径迹探测器用252C f中子源进行剂量刻度,进而确定加速器污染中子辐射剂量。结果与结论:CR-39固体核径迹探测器的刻度系数为(3.76±0.31)×10-5cGy/(径迹数.cm2),加速器污染中子辐射量与主束辐射量的比值在0.05%以内,满足国家职业卫生标准规定的要求。 Objective:To measure neutron contamination in medical electron accelerators using CR-39 solid state nuclear track detectors. Methods:CR-39 solid state nuclear track detectors were calibrated with a standard ^252Cf neutron source. The neutron contamination in the medical electron accelerator was measured by CR-39 detectors. Results and Conclusion: The calibration coefficient of the detector was (3.76 ± 0.31 ) × 10^-5 cGy/( track number · cm^2). The neutron dose was below 0.05% of the main X - ray beam and it fulfilled the requirement of national occupational health standards.
出处 《军事医学科学院院刊》 CSCD 北大核心 2006年第4期346-347,368,共3页 Bulletin of the Academy of Military Medical Sciences
关键词 CR-39固体核径迹探测器 刻度系数 径迹密度 粒子加速器 辐射剂量 CR-39 solid state nuclear track detector calibration coefficient track density particle accelerators radiation dosage