期刊文献+

Research on Electric Impedance Spectroscopy of Living Cell Suspensions by a Chip with Microelectrodes 被引量:1

Research on Electric Impedance Spectroscopy of Living Cell Suspensions by a Chip with Microelectrodes
下载PDF
导出
摘要 A microfabricated electrical impedance spectroscopy (EIS) chip with microelectrodes was developed.The substrate and the electrodes of the chip were made of glass and gold,respectively.The experimental results demonstrated that the EIS-chip could distinguish different solutions (physiological saline,culture medium,living cell suspension etc.) by scanning from 10Hz to 45kHz.A 6-element circuit model was used for fitting the real part and the imaginary part admittance curves of the living cell suspension.An actual circuit was also built and tested to verify the 6-element circuit model proposed.The micro-EIS chip has several advantages including the use of small sample volumes,high resolution and ease of operation.It shows good application prospects in the areas of cellular electrophysioiogy,drug screening and bio-sensors etc. A microfabricated electrical impedance spectroscopy (EIS) chip with microelectrodes was developed. The substrate and the electrodes of the chip were made of glass and gold, respectively. The experimental results demonstrated that the EIS-chip could distinguish different solutions (physiological saline, culture medium, living cell suspension etc.) by scanning from 10Hz to 45kHz. A 6-element circuit model was used for fitting the real part and the imaginary part admittance curves of the living cell suspension. An actual circuit was also built and tested to verify the 6-element circuit model proposed. The micro-EIS chip has several advantages including the use of small sample volumes, high resolution and ease of operation. It shows good application prospects in the areas of cellular electrophysiology, drug screening and bio-sensors etc.
出处 《稀有金属材料与工程》 SCIE EI CAS CSCD 北大核心 2006年第A03期429-432,共4页 Rare Metal Materials and Engineering
关键词 electrical impedance spectroscopy(EIS) MICROELECTRODE living cell suspension 6-element electrical impedance spectroscopy (EIS) microelectrode living cell suspension 6-element
  • 相关文献

参考文献11

  • 1Coster H G L, Chilcott T C, Coster A C F. Bioelectrochemistry and Bioenergetics[J], 1996, 40:79.
  • 2Ayliffe H E, Frazier A B, Rabbitt R D. Journal of Microelectromechanical Systems[J], 1999, 8(1): 50.
  • 3Gomez R, Bashir R, Sarikaya A et al. Biomedical Microdevices [J], 2001, 3(3): 201.
  • 4Cho Y H, Takama N, Yamamoto T et al. Transducers'05[C], Seoul, 2005, 2:1692.
  • 5Lin Y C, Li M, Wu C Y et al. 7th International Conference on Miniaturized Chemical and Biochemical Analysis Systems[C], California, 2003:729.
  • 6Yong Huang, Naveep S Sekhon, James Borninski et al. Sensors and Actuators[J], 2003, 105:31.
  • 7Gowrishankar T R, Weaver J C. Biophysics[J], 2003, 100(6): 3203.
  • 8Huang Y, Rubinsky B. Sensors and Actuators[J], 2001, 89:242.
  • 9Kanai H, Haeno M, Sakamoto K. Medical Progress Technology [J], 1987, 12(2): 1592170.
  • 10Goovaertsyk H G, Faesy Th J C, de G W Valk-de Rooz et al. Physiological Measurement[J], 1998, 19:517.