期刊文献+

方波、差分脉冲伏安法测定水中的镉的最佳条件的研究 被引量:1

Research on the best condition of determining the concentration of cadmium by Square Wave and Differential Pulse Voltammetry
下载PDF
导出
摘要 本实验应用MEC-12B型多功能微机电化学分析仪,采用方波电压伏安法和差分脉冲电压伏安法,以悬汞电极为工作电极,对测定水样中的镉选用何种底液以及多少通氮时间、静置时间、扫描电压有较好的线性结果进行研究.结果表明:在0.01moL/LKI和0.01moL/LNH3-NH4Cl以1:1混合溶液作底液,通氮5分钟,方波电压伏安法静置2分钟,扫描电压为-0.5v~0.1v时,差分脉冲电压伏安法静置4分钟,扫描电压为-0.6v~0.2v时,效果最好.在此条件下Cd2+(10-11–10-9moL/L)之间与峰电流有较好的线性关系. The paper :introduces the condition: of determining the concentration of cadmium by the model MEC-12B Multi-function Microcomputer Electrochemical Analyzer, which uses hanging mercury electrode as work electrode and the method of Square Wave Voltammetry and Differential Pulse Voltammetry: The result is in the mixture of potassiumiodine and Ammonia-Ammonia chcoride buffer solution whose concentration is 0.01molarity, and we should let Nitrogen through the mixture for 5 minutes and keep the mixture still for 2 minutes or 4 minutes, and the concentration of cadmiumion from 10^-11-10^-9 molarity with the peak circulation fit a better line.
出处 《西南民族大学学报(自然科学版)》 CAS 2007年第2期350-355,共6页 Journal of Southwest Minzu University(Natural Science Edition)
关键词 最佳条件 微机电化学分析系统 伏安法 MEC-12B Multi-function Microcomputer Electrochemical Analyzer Cadmium the best condition voltammetrymethod
  • 相关文献

参考文献6

二级参考文献6

共引文献44

同被引文献11

  • 1Zhang W,Lin S,Wang C,et al.PMMA/PDMS valves and pumps for disposable microfluidics[J].Lab on a Chip,2009,9(21):3088-3094.
  • 2Brown L,Koerner T,Horton J H,et al.Fabrication and characterization of poly(methylmethacrylate)microfluidic devices bonded using surface modifications and solvents[J].Lab on a Chip,2006,6(1):66-73.
  • 3Wilson M E,Kota N,Kim Y T,et al.Fabrication of circular microfluidic channels by combining mechanical micromilling and soft lithography[J].Lab on a Chip,2011,11(8):1550-1555.
  • 4Wang W,Zhao S,Pan T.Lab-on-a-print:from a single polymer film to three-dimensional integrated microfluidics[J].Lab on a Chip,2009,9(8):1133-1137.
  • 5Zhang W,Lin S,Wang C,et al.PMMA/PDMS valves and pumps for disposable microfluidics[J].Lab on a Chip,2009,9(21):3088-3094.
  • 6Mark D,Haeberle S,Roth G,et al.Microfluidic lab-on-achip platforms:requirements,characteristics and applications[J].Chemical Society Reviews,2010,39(3):1153-1182.
  • 7Brown L,Koerner T,Horton J H,et al.Fabrication and characterization of poly(methylmethacrylate)microfluidic devicesbonded using surface modifications and solvents[J].Lab on a Chip,2006,6(1):66-73.
  • 8Metz S,Jiguet S,Bertsch A,et al.Polyimide and SU-8 microfluidic devices manufactured by heat-depolymerizable sacrificial material technique[J].Lab on a Chip,2004,4(2):114-120.
  • 9Pan T,Wang W.From cleanroom to desktop:emerging micro-nanofabrication technology for biomedical applications[J].Annals of biomedical engineering,2011,39(2):600-620.
  • 10余若祯,王红梅,方征,闫振广,王宏.重金属离子快速检测技术研究与应用进展[J].环境工程技术学报,2011,1(5):438-442. 被引量:31