摘要
多元光谱拟合(MSF)法是一种新的校正多种基体干扰的方法.但应用于测定生物样品中痕量元素的分析较少,在猪肝样品的痕量分析中未见报道,这是由于猪肝样品中基体成分复杂的原因.本文采用MSF法,有效地校正了复杂基体的光谱干扰,测定了猪肝样品中痕量的Pb、Mo,得到满意的测定结果,测得Pb、Mo两种元素的检出限分别为5.02μg.L-1、2.32μg.L-1,方法的回收率分别为97.2%、92.1%,方法的精密度分别为1.22%、1.35%.
Multicomponent spectral fitting(MSF) is an important method of interference correction. It is little used for analysis of biological samples because of the complicated matrix. In this thesis, MSF method is used to correct the complicated matrix interference. Trace element in pork liver had been determinated using this method, got satisfied results. The detection limts of Pb and Mo were 5.02μg/L, 2.32μg/L. The recoveries of Pb and Mo were 97.2% ,92.1%. The relative standard deviations of Pb and Mo were 1.22%, 1.35%.
出处
《华中师范大学学报(自然科学版)》
CAS
CSCD
2007年第4期557-560,共4页
Journal of Central China Normal University:Natural Sciences
基金
国家自然科学基金资助项目(20603010).
关键词
端视ICP
基体效应
多元光谱拟合
校正原理
axially viewed inductively coupled plasma
matrix effects
multicomponent spectral fitting( MSF)
correction principle