摘要 以固定化胆碱氧化酶(EC 5896)复合H2O2电极构成电流型酶电极生物传感分析仪,酶电极测定线性范围:0 mg·L^-1~200 mg·L^-1,精度RSD〈1.5%,响应时间:40 s,固定化胆碱氧化酶膜使用寿命大于60 d,实际测定氯化琥珀胆碱注射液中氯化胆碱含量回收率:100.3%~102.3%,采用酶电极法测定氯化琥珀胆碱注射液中氯化胆碱含量具有专一性高、成本低、分析速度快等优点。 Choline oxidase was immobilized at hydrogen peroxide electrode and the enzyme electrode was used for the amperometric determination of choline chloride. A good linear scope was from 0 to 200 mg·L^-1, the corresponding relative standard deviation less than 1.5%. The response time for enzyme electrode with choline was 40 seconds, the immobilized enzyme membrane lasted at least 60 days, and the recovery mtewas from 100.3 % to 102.3 %. The contents of choline chloride in succinyl choline chloride injection can be determined specially, rapidly and inexpensively with enzyme electrode.
出处 《山东科学》 CAS 2007年第6期22-24,共3页 Shandong Science
关键词 酶电极 胆碱氧化酶 胆碱测定 氯化琥珀胆碱注射液 enzyme electrode choline oxidase choline analysis succinyl choline chloride injection
  • 相关文献

参考文献4

  • 1中国药典委员会.中华人民共和国药典二部[M].北京:化学工业出版社,1990:附录817-818.
  • 2冯德荣,冯云水,周万里,尚雪芹,张冬雁.固定化乳酸酶膜圈的研究[J].山东科学,1990,3(3):1-6. 被引量:12
  • 3LUIGI C,et al. The Hydrogen Peroxide Sensing Plantiun Anodes as an Analytical Enezyme Electrode[J]. Clinic Chimica Acta, 1985, 151 : 71 - 83.
  • 4PANDEY P C, LANTREIBECQ F. Study on choline Chloride by Enzyme Electrode[J]. Applied Biochem and Biotech, 1991,31:145.

二级参考文献2

  • 1冯炜权,冯美云.乳酸与运动能力 一、运动时乳酸生成和肌肉、血液乳酸[J]中国运动医学杂志,1987(03).
  • 2杨奎生,王世平.用0.02毫升全血分别测定血乳酸和血糖的超微量方法[J]中国运动医学杂志,1983(02).