期刊文献+

自建生化检测系统的性能评价 被引量:6

Estimation on the capability of a self-developed biochemical detection system
下载PDF
导出
摘要 目的探讨不同检测系统间实验结果的可比性,以证实自建检测系统的试验结果的可靠性。方法按美国CLIA'88能力比对检验的分析要求,比对实验应建立在检测系统的分析性能的基础上。本文分析性能评价包括不精密度,不准确度、线性,分析灵敏度(检测线),方法学比较。结果自建系统项目丙氨酸氨基转移酶(ALT)、尿素(Urea)、肌酐(Crea)、总胆固醇(TC)、血清蛋白A1(ApoA1)、人血清脂蛋白B(ApoB)、高密度脂蛋白胆固醇(HDL-C)与Roche可溯源参考检测系统试验结果相关系数>0.975。低密度脂蛋白胆固醇(LDL-C)存在方法学的差异试验结果相关系数<0.975。检测不精密度CV值满足美国CLIA'88能力比对检验的质量分析要求。线性检测范围满足临床检测线性要求,特别线性高值要求。结论自建检测系统性能评价试验结果与Roche可溯源参考检测系统试验结果比对具有相关性,不确定度也相应增加。各种性能相关性满足统计学原理并不证明其具有溯源性。 Objective To explore the comparability of results among different detection systems so as to prove the result reliability of self-developed biochemical detection system. Methods Proficiency testing should be based on performance analysis of detection system meeting to the demand of American CLIA88 proficiency testing, The performance estimation consisted of imprecision, inaccuracy, linearity, sensitivity of measurement (detection limit) and methodical comparison. Results There were no ststistical differences of results of alanine aminotransferase (ALT), creatine (Cr), total cholesterol (TC), apolipoprotein A1 (ApoA1), apolipoprotein B (ApoB), as well as high-density lipoprotein cholesterol (HDL-C) detected by self-developed biochemical detection system and by traceable Roche Modular reference assay system. Their respective relative coeffieients were all above 0. 975, while there was significant difference of methodical analysis for low-density lipoprotein cholesterol (LDL-C) between two systems, and the relative coeffieient was below 0. 975. The CV of imprecision of measurement met the demand of American CLIA88 proficiency testing quality control analysis. The range of linear detection met the demand of clinical detection, eapecially for linearity of high value. Conclusion There is relativity of detected results by self-developed biochemical detection system and by traceable Roche Modular reference assay system, and the uncertainty increases accordingly. Traceability can't be proved only by significant performance relativity.
出处 《检验医学与临床》 CAS 2008年第1期5-8,共4页 Laboratory Medicine and Clinic
基金 江苏省徐州市科技局社会发展项目(XZ2006238)
关键词 生化检测系统 可比性 溯源性 biochemical detection system comparison traceability
  • 相关文献

参考文献3

二级参考文献3

  • 1ICSH.Reference method for the enumeration of erythrocytes and leucocytes, Chin Lab Heamat, 1994,16:131-138.
  • 2ICSH.The assignment of values to fresh blood used for calibration automated blood cell counters. Clin Lab Haemat, 1988, 10:203-212.
  • 3彭明婷,申子瑜.血细胞分析溯源体系的建立及有关问题的探讨[J].中华检验医学杂志,2004,27(3):132-133. 被引量:67

共引文献49

同被引文献44

引证文献6

二级引证文献25

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;