摘要
在采用美国试验与材料学会(ASTM)D6591标准方法测定柴油芳烃含量的过程中,通过光电二极管阵列检测器(PDAD)检测得到的色谱保留时间值及光谱特征吸收曲线,观察柴油中各类芳烃在极性氨基柱上的分离情况;用PDAD检测器考察了根据ASTMD6591确定的反冲洗时间点B对分析结果的影响,结果发现其允许的变化范围为B±0.2min。在单环芳烃和双环芳烃分界不明显时,用PDAD检测器可以确定其切割点。
When using ASTM D6591 to determine the content of aromatic hydrocarbons in diesel oil,a photodiode array detector(PDAD)was used to monitor the separation of aromatic hydrocarbons.The influence of back flush time B was studied and it was observed the results of di-aromatic hydrocarbons or poly-aromatic hydrocarbons will give errors when the back flush time is more than B±0.2 min.The PDAD can also be used to determine the best cutting time between mono-aromatic hydrocarbons and di-aromatic hydrocarbons,when the separation was not baseline resolved.
出处
《色谱》
CAS
CSCD
北大核心
2008年第2期250-253,共4页
Chinese Journal of Chromatography
关键词
高效液相色谱
光电二极管阵列检测
芳烃类型
柴油
high performance liquid chromatography(HPLC)
photodiode array detection
aromatic hydrocarbon types
diesel oil