摘要 根据透镜成像规律和单模光纤的传输特性,获得基于单模光纤的共聚焦内窥成像系统轴向光强分布,发现系统的成像过程由于单模光纤的介入呈现非线性,照明光路系统和接收光路系统仅影响系统接收信号的总强度,系统层析能力取决于探头和单模光纤性能.通过建立的共聚焦内窥成像实验装置,测量出在三组不同物镜组合情况下系统的归一化轴向光强分布.结果显示已建立的共聚焦内窥成像系统的理论模型是正确的,扫描光点的定位精度对系统层析能力影响很大,系统设计应该综合考虑光学、机械和电子方面的性能要求. Based on image theory of lens and transmission characteristic of single-mode fiber (SMF), axial intensity distribution of confocal endoscopic imaging system based on SMF is presented. It is found that imaging by the system is nonlinear duing to SMF,illumination system and reception system both affect the sum optical signal intensity,and optical sectioning strength of the system is depended on prober and SFM. Using the experimental set-up of confocal endoscopic image, normalized axial intensity distributions under three different kinds of convertible objective are obtained. The results show that the presented theoretical model of confocal endoscopic imaging system is correct, location precision of scanning point also affect optical sectioning strength of the system, and its design should be allowed for synthetically the capability of optics,mechanism and electron.
作者 刘勇 陈家璧
出处 《光子学报》 EI CAS CSCD 北大核心 2008年第6期1152-1155,共4页 Acta Photonica Sinica
基金 973国家重点基础研究发展规划(2005CB724304) 上海市科委光技术专项(04dz05114)资助
关键词 共聚焦内窥成像 层析能力 单模光纤 探头 Confocal endoscopic image Optical sectioning strength Single-mode fiber Prober
  • 相关文献

参考文献8

  • 1杨初平,旷卫民,刘军,唐志列.二次谐波共焦成像的分辨率[J].光子学报,2006,35(11):1709-1712. 被引量:5
  • 2葛华勇,王保华,任秋实.共聚焦成像系统中探测器对分辨率的影响(英文)[J].光子学报,2006,35(4):587-590. 被引量:12
  • 3刘勇,陈家璧.非相干光对共聚焦显微镜层析能力的影响[J].光子学报,2007,36(2):316-319. 被引量:1
  • 4GMITRO A F,AZIZ D. Confocal microscopy through a fiberoptic imaging bundle [J]. Opt Lett, 1993,18(8) : 565-567.
  • 5KNITTEL J, SCHNIEDER L, BUESS G, et al. Endoscope-compatible confocal microscope using a gradient index-lens system[J]. Opt Comm, 2001,188 (2) : 267-273.
  • 6LIANG C,DESCOUR M R,SUNG K B,et al. Fiber confocal reflectance microscope (FCRM) for in-vivo imaging [J]. Opt Express ,2001,9(13): 821-830.
  • 7吴重庆.光薄导理论[M].北京:清华大学出版社,2000:108.
  • 8GU Min,SHEPPARD C J R,Gan X.. Image formation in a fiber-optical confocal scanning microscope [J]. JOSA A, 1991,8(11):1755-1761.

二级参考文献22

  • 1Moreaux L,Sandre O,Blanchard-Desce M,et al.Membrane imaging by simultaneous second-harmonic generation and two-photon microscopy.Optics Letters,2000,25(5):320~322
  • 2Guo Yici,Ho P P,Savage H,et al.Second-harmonic tomography of tissue.Optics Letters,1997,22(17):1323~1325
  • 3Alvin T Yeh,Nader Nassif,Aikaterini Zoumi,et al.Selective corneal imaging using combined second-harmonic generation and two-photon excited fluorescence.Optics Letters,2002,27(23):2082~2084
  • 4Malcom Brown R,Jr,C Millard Andrew,et al.Macromolecular structure of cellulose studied by second-harmonic generation imaging microscopy.Optics Letters,2003,28(22):2207~2209
  • 5Jiang Yi,Ivan Tomov,Wang Yimin,et al.Second-harmonic optical coherence tomography.Optics Letters,2004,29(10):1090~1092
  • 6Fu Ling,Gan Xiaosong,Gu Min.Use of a single-mode fiber coupler for second-harmonic-generation microscopy.Optics Letters,2005,30(4):385~387
  • 7Minoru Kobayashi,Katsumasa Fujita,Tomoyuki Kaneko,et al.Second-harmonic-generation microscope with a microlens array scanner.Optics Letters,2002,27(15):1324~1326
  • 8Shen Y R.The Principles on Nonlinear Optics,New York:John Wiley & Sons Inc.1984.71~84
  • 9Goodman J W.Introduction to Fourier Optics,New York:John Wiley & Sons Inc.1968.88~155
  • 10WILSON T,CARLINI A R.Three-dimensional imaging in confocal imaging system with finited sized detectors[J].J Microscopy,1988,149(1):51-66.