期刊文献+

金汞齐富集冷原子荧光测汞法的研究 被引量:5

Datermination of Meacury by Non-Dispersive Atomic Fluorescence Spectrometer With Gold Amalgamation Concentration
下载PDF
导出
摘要 本文介绍了一种用于测定汞的金汞齐富集热解系统。研究了该系统与冷原子荧光测汞仪联用的工作条件和干扰,在最佳条件下的检出下限为4×10^(-13)克汞,分析周期3分钟,相对标准偏差2.5%(1.5×10^(-9)克汞)。 A non-dispersive atomic fluorescence system determining mercury which makesuse of Au-coated chromatographic supports for collection of Hg and a water-coolingelectric furnace for desorption of Hg are described. The experimental conditions andthe effects of some organic solvents vapour on determination of Hg have been investi-gated. After optimization the detection limit of 0.0004 ng of Hg, the relative standarddeviation of 2.5%(1.5 ng of Hg, n=10) and the cycle time of 3 min. for entireprocese are obtained. Results are reported for application of the system to a varietyof samples, including gases, liquids and solids.
出处 《分析测试通报》 CAS CSCD 1989年第2期29-32,共4页
关键词 金汞齐 富集 冷原子荧光法 热解 Mercury gold amalglamation non-dispersive atomic fluoresence spectrometry
  • 相关文献