摘要 目的观察Sysmex XS800i血液分析仪提示白细胞异常的可信程度。方法对血液分析仪提示白细胞异常信号的标本同时进行显微镜镜检,并对比分析。结果Sysmex XS800i血液分析仪对正常的成熟白细胞分类已达到较理想的程度,对异型淋巴细胞和原始、幼稚细胞的提示,与显微镜镜检结果符合程度低于成熟细胞。结论血液分析仪不能完全替代显微镜镜检,一旦仪器有异常细胞警告提示或临床高度怀疑细胞形态异常时,必须作涂片显微镜镜检。 Objective To study the confidence level of indicating abnormal leukocyte by Sysmex-XS800i blood analyzer. Methods When specimens were indicated by Sysmex-XS800i blood analyzer as abnormality,they were contrastively inspected by microscope simuhaneously. Results Using Sysmex-XS800i blood analyzer on classifying normal maturity leukocyte was considered ideal, but was unfit for abnormal lymphocyte such as primitive and immature cells when compared with microscope test. Conclusion Blood analyzer couldn't replace microscope test completely. When abnormal cells were indicated by blood analyzer or suspected by clinician,these specimens must be smeared for microscope test.
出处 《临床输血与检验》 CAS 2008年第4期300-302,共3页 Journal of Clinical Transfusion and Laboratory Medicine
关键词 SYSMEX XS800i血液分析仪 白细胞异常 显微镜镜检 Sysmex-XS800i blood analyzer Abnormal leukocyte Microscope test
  • 相关文献

参考文献3

  • 1Herklote R,Huber AR. Precision and accuracy of the leukocyte differential on the XE-2100 [J]. Sysmex J Int, 2001,11 (1) : 8-21.
  • 2中华医学检验学会临床血液和尿液分析专题研讨会纪要.中华医学检验杂志,1995,(2):125-125.
  • 3陈小剑,王晓欧,李绵绵,黄晓彤,朱丽丹.XE-2100血细胞分析仪血涂片复检标准制定及评价[J].中华检验医学杂志,2007,30(4):384-387. 被引量:29

二级参考文献6

  • 1Grimaldi E, Scopacasa F. Evaluation of the Abbott CELL-DYN 4000 hematology anlyzer. Am J Clin Pathol,2000,113:497-505.
  • 2Amourouxi, Balay M, Marfaing KA. Evaluation of the blood analyzer Beckman Coulter LH750 : analytic performance, decision rules. Ann Biol Clin(Paris) ,2003,61:576-584.
  • 3Kunicka JE, Fischer G, Murphy J, et al. Improved platelet counting using two-dimensional laser lingts scatter. Am J Clin Pathol,2000,114 :283-289.
  • 4National Committee for Clinical Laboratory Standards. Reference leukocyte differential count (proportional) and evaluation of instrument methods: approved standard. NCCLS Document H20-A[S]. NCCLS, Wayne(USA), 1992.
  • 5倪宗錾,主编.卫生统计学.4版.北京:人民卫生出版社,2002:165-167.
  • 6Herldotz R, Huber AR. Precision and accuracy of the leukocyte differential in the sysmex XE-2100. Sysmex J Int,2001,11:8-21.

共引文献28

同被引文献27

引证文献5

二级引证文献9

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部