OrbscanⅡ系统与A超角膜测厚仪测量近视眼角膜中央厚度的比较 被引量:6

Comparison of central corneal thicknesses measured by Orbscan Ⅱ system and Ultrasonic pachymetry in myopia
下载PDF
导出
摘要 目的将orbscanⅡ系统与A超角膜测厚仪测量的近视眼中央角膜厚度的结果,进行对比分析,为临床手术提供准确的数据。方法选取近视患者200例(400只眼),按屈光度、A超测得角膜厚度不同分成4组。分别应用OrbscanⅡ系统及A超角膜测厚仪对角膜中央厚度进行测量。结果OrbscanⅡ系统与A超角膜测厚仪测得角膜中央厚度分别为(568.06±34.29)/μm和(536.25±33.90)μm,两者测量结果存在差异(P〈0.01)。OrbscanⅡ系统测量结果较A超角膜测厚仪结果厚(31.81±16.39)μm。按屈光度不同分组,即低度数组-0.50- -3.00D,中度数组-3.25~-6.00D,高度数组-6.25- -9.00D,超高度数组-9.25~-15.00D,OrbseanⅡ系统与A超角膜测厚仪两者之间差异显著(P〈0.01)。按A超测得角膜厚度分为4组,即I组≤500μm,Ⅱ组500~536μm,Ⅲ组536—572μm,Ⅳ组≥572μm。OrbscanⅡ系统与A超角膜测厚仪两者之间差异显著(P〈0.01)。结论OrbscanⅡ系统中央角膜厚度测量值比A超角膜测厚仪测量值厚约30μm,OrbscanⅡ系统不能完全替代A超角膜测厚仪。 Objective To evaluate the corneal thinknesses measured with Orbscan Ⅱ system and A - ultrasonic pachymetry. Two hundred myopic subjects(400eyes) were recruited and divided into four groups according to the diopter and corneal thickness measttred by A- ultrasonic pachymetry. Central corneal thicknesses (CCT) were measured with Orbscan Ⅱ topography system and A - ultrasonic pachymetry.Results The mean CCTby Orbscan Ⅱ system and A- ultrasonic pachymetrywere (568.06±34.29) μm and (536.25±33.90)μm respertively.The result of Orbscan Ⅱ system was higher than that of A- ultrasonic pachymetry with a difference of (31.81 ± 16.39) μm ( P 〈 0.01 ). Firstly, The subjects were divided into four groups according to diopter: low myopia( -0.50 - - 3.00D ), medium myopia( - 3.25- - 6.00D) ,high myopia( - 6.25 - - 9.00D)and supermyopia( - 9.25 - - 15.00D), The results of Orbscan Ⅱ system were higher than those of A - ultrasonic pachymetry respectively( P 〈 0.01 ). Secondly, The subjects were divided into four groups based on the CCT measured by A - ultrasonic pachymetry : group1( ≤ 500 μm), group 2 (500- 536 μm), group 3 (536 - 572μm), and group 4( ≥ 572 μm). The results of Orbscan Ⅱ system were higher than those of A - ultrasonic pachymetry ( P 〈 0.01 ). Conclusions There were significant differencees between the results of these two apparatus. The outcomes of Orbscan Ⅱ system are approximately 30μm higher than those of A- ultrasonic pachymetry. Orbscan Ⅱ system cannot totally replace A - ultrasonic pachymetry.
出处 《武警医学》 CAS 2008年第10期880-882,共3页 Medical Journal of the Chinese People's Armed Police Force
关键词 OrbscanⅡ A超角膜测厚仪 中央角膜厚度 Orbscan Ⅱ system Ultrasonic pachymetry Central corneal thickness
  • 相关文献

参考文献9

二级参考文献32

  • 1李镜海,周芳,周树安,李树秋,刘石磷,巫洁.正常眼与近视眼多方位角膜厚度研究[J].中华眼科杂志,1994,30(6):445-448. 被引量:32
  • 2李凤鸣.眼科全书(上册)[M].北京:人民卫生出版社,1997.842.
  • 3Nader G,Iskander,Ellen Anderson Penno.Accuracy of Orbscan pachymetry measurements and DHG ultrasound pachymetry in primary laser in situ keratomileusis and LASIK enhancement procedures[J].J Cataract Refract Surg,2001,27(5):681-685.
  • 4Liu ZG,Andrew J,Pflugfelder SC,et al.Evaluation of cornealthickness and topography in normal eyes using the Orbscan cornealtopography system[J].Br J Ophthalmol,1999,83(8)∶774-778.
  • 5Yaylali V,Kaufman SC,Thompson HW.Corneal thickness measurements with the Orbscan topography system and ultrasonic pachymetry[J].J Cataract Refract Surg,1997,23(7):1345-1350.
  • 6Nissen J,Hjortdal J,Ehlers N,et al.A clinical comparisonof optical and ultrasonic pachometry[J].Acta Ophthalmol(Copenh),1991,69(4):659-663.
  • 7Seiler T,Koufala K,Gichter G.Iatrogenic keratectasiaafter laser in situ keratomileusis[J].J Refract Surg,1998,14(4):312-317.
  • 8Joo CK,Kim TG.Corneal ectasia detected after laser insitu keratomileusis for correction of less than 12 diopters of myopia[J].J Cataract Refract Surg,2000,26(2):292-295.
  • 9李凤鸣,眼科全书.上,1997年,842页
  • 10Baek T M,Ophthalmology,2001年,108卷,317页

共引文献50

同被引文献57

引证文献6

二级引证文献20

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部