取样对激光法测定导热系数的影响 被引量:2

Effect of sampling on thermal conductivity determination by laser flash method
下载PDF
导出
摘要 以建筑陶瓷、石墨砖和Si3N4-SiC砖为主要试样,研究取样数量、取样位置及其方向对激光法测定材料导热系数结果的影响。结果表明,对于均质材料,随着取样数量的增加,导热系数的测定结果趋于其真值,结合目前设备的技术条件,认为取2~3个样品进行测试即可获得较理想的结果;而对于非均质材料,取样位置和方向都对其导热系数的测定结果有很大的影响,这主要是由于非均质材料的结构或组成随着取样位置和方向的不同而造成的,非均质材料的取样数量至少为5个。 Effect of sampling amount, position and direction on thermal conductivity determination by laser flash method was studied by using building ceramics, graphite bricks and Si3 N4-SiC bricks as experimental materials. The results show that: for homogeneous materials, determined thermal conductivity moves close to the true value with the sample number increasing, and with contemporary technical conditions 2-3 samples are needed for satisfactory results; for heterogeneous materials, sampling position and direction have serious influence on thermal conductivity determination, which can be explained by different compositions and structures caused by the un-uniformity of materials, and therefore no fewer than 5 samples are needed.
出处 《武汉科技大学学报》 CAS 2009年第2期193-196,204,共5页 Journal of Wuhan University of Science and Technology
关键词 取样 导热系数 建筑陶瓷 石墨砖 Si3N4-SiC砖 sampling thermal conductivity building ceramics graphite brick Si3N4-SiC brick
  • 相关文献

参考文献10

  • 1孙晓刚,戴景民,丛大成,褚载祥.材料热物性脉冲加热测量技术[J].仪器仪表学报,2002,23(z1):341-342. 被引量:5
  • 2葛山,尹玉成.激光闪光法测定耐火材料导热系数的原理与方法[J].理化检验(物理分册),2008,44(2):75-78. 被引量:20
  • 3全国耐火材料标准委员会,中国标准出版社第五编辑室.耐火材料标准汇编上册[S].北京:中国标准出版社,2007:505-562.
  • 4冶金工业信息标准研究院标准化研究所,中国标准出版社第二编辑室.炭素制品及其试验方法标准汇编[S].北京:中国标准出版社,2002:105-117.
  • 5Parker W J, Jenkins R J, Butler C P. Flash method of determining thermal diffusivity, heat capacity and thermal conductivity[J]. Journal of Applied Physics, 1961, 32(9): 1 679-1 684.
  • 6Righini F, Cezairliyan A. Pulse method of thermal diffusivity measurements[J]. High Temperatures- High Pressures, 1973(5): 481-501.
  • 7奚同庚 顾龙文 顾宗义 等.激光闪光法导热系数测定仪的研究.新型无机材料,1974,3(3):37-37.
  • 8奚同庚 周新宇 李宗杰 等.计算机运控的激光导热仪的研究.工程热物理学报,1980,1(2):147-155.
  • 9何冠虎 叶庆和 董淑清 等.同时测量导温系数和比热容的激光加热降温法.计量学报,1983,4(3):168-168.
  • 10ASTM E1461-2001, Standard test method for thermal diffusivity by the flash method[S].

二级参考文献16

  • 1孙晓刚,戴景民,丛大成,褚载祥.材料热物性脉冲加热测量技术[J].仪器仪表学报,2002,23(z1):341-342. 被引量:5
  • 2张亚静,孙萍.含碳(或碳化硅)耐火材料热导率的测试[J].耐火材料,2007,41(2):156-158. 被引量:7
  • 3[1]F.Righini.System for fast high-temperature measurements.High Temp High Press,1972,4(4):597~603.
  • 4[2]F.Righini.Normal spectral emissivity of Niobium by a pulse-heating reflectometric technique.Int J Thermophys,1999,20(4):1095~1105.
  • 5[3]A.Cezairliyan.Simultaneous measurements of normal spectral emissivity by spectral radiametry and laser polarimetry.Int J Thermophys,1996,17(6):1455~1473.
  • 6全国耐火材料标准委员会,中国标准出版社第五编辑室.耐火材料标准汇编上册[M].第3版.北京:中国标准出版社,2007:505-562.
  • 7冶金工业信息标准研究院标准化研究所,中国标准出版社第二编辑室.炭素制品及其试验方法标准汇编[M].第2版.北京:中国标准出版社,2002:105-117.
  • 8Parker W J, Jenkins R J, Butler C P. Flash method of determining thermal diffusivity, heat capacity and thermal conductivity[J]. Journal of Applied Physics, 1961,32(9) : 1679-1684.
  • 9Righini F, Cezairliyan A. Pulse method of thermal diffusivity measurements[J]. High Temperatures-High Pressures, 1973,5(4) :481-501.
  • 10奚同庚 顾龙文 顾宗义 等.激光闪光法导热系数测定仪的研究.新型无机材料,1974,3(3):37-37.

共引文献23

同被引文献21

引证文献2

二级引证文献2

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部