期刊文献+

微型电化学系统中的微电流测量 被引量:5

Weak-current Measurement in a Micro-electrochemistry System
下载PDF
导出
摘要 介绍了一个微型电化学系统中的微电流测量技术.在电路面积和元件精度受限的情况下,通过电流-电压转换电路的设计,实现了高灵敏度和大量程.通过单片机中的数字校准,补偿了模拟元件带来的误差.该系统对于微电流测量的分辨率达到pA量级,精度达到0.1%,是一套适合工业应用的高集成度和高精度的系统. The paper introduces the weak-current measurement technology in a micro-electrochemistry system. In restriction of circuit area and component precision, high sensitivity and large measurement range is implemented by the design of current-voltage transition circuit. Digital calibration in the Micro Controller Unit compensates the error introduced by the analog components. The resolution of current measurement has reached pA level, and the precision has reached 0.1%. With high integrity and precision, this system is suitable for industrial application.
出处 《微电子学与计算机》 CSCD 北大核心 2009年第6期1-4,共4页 Microelectronics & Computer
关键词 电化学系统 微电流测量 数字校准 electrochemistry system weak-current measurement digital calibration
  • 相关文献

参考文献6

二级参考文献3

  • 1张祖训.超微电极电化学[M].北京:科学出版社,1988..
  • 2Microcontroller Handbook. Intel Corp., 1986.
  • 3张祖训,超微电极电化学,1988年

共引文献24

同被引文献39

  • 1MILLER R A. A novel micromachined high-field asymmetricwaveformion mobility spectrometer[J]. Sensor- sand Actuators B Chemical 2000, 67: 300-306.
  • 2GUERVREMONT R. High-field asymmetric waveformion mobility spectrometry: A new tool for mass spectrometry [J]. Journal of Chromatography A, 2004, 1058: 3-19.
  • 3MILLER R A, EICEMAN G A, NAZAROV E G et al. A MEMS radio-frequency ion mobility spectrometer for chemical agent detection[J]. Sensors and Actuators A, 2001, 91: 307-318.
  • 4EICEMAN G A. A micro-machined ion mobility spectrometer-mass spectro-meter[J]. International Journal for Ion Mobility Spectrometry, 2000, 3(1): 15-27.
  • 5MILLER R A, EICEMAN G A, NAZAROV E G. A micromachined field asymmetric ion mobility spectrometer(FA-IMS)[J]. International Journal for Ion Mobility Spectrometry, 2002, 5(2): 118-134.
  • 6WANG L. Low voltage, high speed, high speed, switched capacitor circuit[J]. IEEE Journal of Solid State Circuit, 2003, 38 (8): 1411-1416.
  • 7BORSDORF H, NAZAROV E G.MILLER R A. Atmospheric-pressure ionization studies and field dependence of ion mobilities of isomeric hydrocarbons using a miniature differential mobility spectrometer [J]. Analytica Chimica Acta, 2006, 575(1): 76-88.
  • 8NAZAROV E G MILLER R A, EICEMAN G A, et al.Miniature differential mobility spectrometry using atmospheric pressure photoionization[J]. Anal. Chem., 2006, 78, 4553-4563.
  • 9PAPANASTASIOU D, WOLLNIK H, RICO G et al. Differential mobility separation of ions using a rectangular asymmetric waveform [J]. Journal of Physical Chemistry, 2008, 112(16): 3638-3645.
  • 10MILLER R A. A novel micromachined high-field asymmet- ricwaveform-ion mobility spectrometer [J]. Sensorsand Actuators B Chemical, 2000, 67: 300-306.

引证文献5