基于容积补偿法和电阻抗法的小型循环动态监测系统的开发 被引量:7

Development of a Handy-type Monitoring System for Cardiovascular Haemodynamic Functions Based on the Volume-compensation and Electrical Admittance Method
下载PDF
导出
摘要 介绍了一种基于容积补偿法和电阻抗法的小型循环动态监测系统的开发。系统中,采用指动脉局部加压技术及相应的检测装置,改善了全周加压袖套所存在的严重的淤血问题;运用新型检测方法,提高了信号检测的稳定性;应用移动线圈式电空变换器,大幅降低了系统使用的噪声。研究还探讨了心拍出量连续检测时胸部两重圆筒模型的适用性问题,以及用点状电极替代带状电极时的最佳配置。试验表明,开发的监测系统可以满足人们在自由行动下每一心拍循环动态变化的检测,携带舒适性有了明显提高,可实现了循环动态信息的无约束无创性长时间连续检测。 A handy-type monitoring system for cardiovascular haemodynamic functions based on the volume-compensation and electrical admittance method is developed. In this system, the inconvenient and discomfort due to cuff occlusion of the biological segment for BP measurement is improved by developing a new device using a local pressurization method, the stability for the physiological information detection is improved by developing a new detecting system, and the noise during detection using the system is greatly reduced by using a newly developed nozzle-flapper type electro-pneumatic converter. Also, for electrical admittance cardiography to estimate CO, the applicability of a two- compartment coaxial cylindrical model and the optimal position of a spot-electrode array that is used to replace the conventional band-electrode are discussed in this research. Experimental result shows that the monitoring system should satisfy non-invasive BP and CO measurement on beat by beat, and the comfort of measurement is significantly improved.
出处 《中国医疗器械杂志》 CAS 2009年第3期167-171,共5页 Chinese Journal of Medical Instrumentation
关键词 容积补偿法 电阻抗法 无创性检测 循环动态机能 Volume-compensation method, Electical admittance method,Noninvasive measurement, Cardivoascular haemodynamic
  • 相关文献

参考文献7

  • 1K.Yamakoshi,H.Simazu,M.Shibata,et al.New oscillometric method for indirect measurement of systolic and mean arterial pressure in the human finger Part 1 and Part 2[J].Med & Biol.Eng.& Comput,1982,20:307-318.
  • 2K.Yamakoshi,H.Simazu,M.Shibata et al.Indirect measurement of instantaneous arterial blood pressure in the human finger by the vascular unloading technique[J].IEEE Trans.Biomed.Eng.1980,BME-27:150-155.
  • 3K.Yamakoshi and M.Nakagawara.Voltage clamp method for the use of electrical admittance plethysmography in human body segments[J].Med & Biol.Eng.& Comput,1995,33:740-743.
  • 4H.Shimazu,H.Ito,H.Kobayashi et al.New idea to measure diastolic arterial pressure by the volume-oscillometric method in human fingers[J].Med & Biol.Eng.& Comput,1986,24:549-554.
  • 5J.Nyboer.Electrical impedance plethysmography:A physical and physiological approach to peripheral vascular study[J].Circulation,1950,2:811-821.
  • 6K.Yamakoshi,S.Tanaka and H.Shimazu.Electrical admittance cuff for non-invasive and simultaneous measurement of hematocrit,arterial pressure and elasticity using volume-oscillometric method[J].Med & Biol.Eng.& Comput,1994,32:99-107.
  • 7M.Pagani,G.Lombardi,S.Guzzetti et al.Power spectral analysis of heart rate an arterial pressure variabilities as a marker of sympatho-vagal interaction in man and conscious dog[J].Circulation,1986,59:178-193.

同被引文献51

  • 1陆渭明,李和平,蒋正尧,崔鲁宁,陶上乘,张冬梅,涂继善,缑东元.对利用脉搏波传播速度间接测量血压时主要影响因素的研究[J].中国生物医学工程学报,1995,14(1):89-93. 被引量:14
  • 2向海燕,俞梦孙.用脉搏波传导时间实现血压连续测量[J].医疗卫生装备,2006,27(2):19-21. 被引量:31
  • 3K. Yamakoshi, H. Simazu, M. Shibata, et al. Indirect measurement of instantaneous arterial blood pressure in the human finger by the vascular unloading technique[J]. IEEE Trans. Biomed. Eng, 1980, BME-27, 150-155.
  • 4K. Yamakoshi, H. Simazu, M. Shibata, et al. Indirect measurement of instantaneous arterial blood pressure in the rat[J]. Am. J. Physiol, 1979, 237, H632-H637.
  • 5H. Shimazu, H. Ito, H. Kobayashi, et al. New idea to measure diastolic arterial pressure by the volume-oscillometric method in human fingers[J]. Med & Biol. Eng. & Comput, 1986, 24:549-554.
  • 6S. Tanaka and K. Yamakoshi. Ambulatory instrument for monitoring indirect beat-by-beat blood pressure in superficial temporal artery using volume-compensation method[J]. Med & Biol. Eng. & Comput, 1996, 34, 441-447.
  • 7S. Tanaka, S. Gao, M. Nogawa, et al. Non-invasive Measurement of Instantaneous Blood pressure in the Radial Artery[J]. IEEE Engineering in Medicine and Biology Magazine, 2005, 24(4): 32-37.
  • 8NYBOER J. Electrical impedance plethysmogrsphy: A physi- cal and physiological approach to peripheral vascular study [J]. Circulation, 1950, 2:811-821.
  • 9KUBICEK W G, PATERSON R P, WITSOE D A. Impedance cardiography as a noninvasive method of monitoring cardiac function and other parameters of the cardiovascular system [J]. Ann NY Acad Sci, 1970, 170; 724-732.
  • 10PENNEY B C, PATWARDHAN N A, WHEELER H B. Simplified electrode array for impedance cardiography [J]. Med Biol Eng & Comput, 1985, 23,1-7.

引证文献7

二级引证文献13

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部