成人颅脑CT低剂量平扫成像的临床初步研究 被引量:23

Preliminary Clinical Study of Dose-reducing Protocols in Adult Cranial CT
下载PDF
导出
摘要 目的:评价成人颅脑低剂量CT平扫的图像质量及临床价值。材料和方法:110例患者被随机分组并运用64层螺旋CT机采用非螺旋扫描方式分别行330mAs、130mAs、100mAs的颅脑平扫,对相应图像进行主观评分、噪声测量和客观评价分析。结果:130mAs和100mAs组小脑和内囊评分均低于侧裂池评分;各组间各相应解剖结构评分均具有统计学差异(P<0.05)。各组内小脑白质噪声大于壳核和半卵圆中心噪声,随着管电流量的降低,各相应层面各解剖结构噪声随之增加(P<0.05)。颅脑平扫CT图像的对比噪声比则随着管电流量的降低而减少(P<0.05),而灰质清晰度无明显变化。相对于330mAs组,130mAs组DLP减少了61%,100mAs组减少了70%。结论:130mAs颅脑平扫CT图像质量可以常规用于成人临床,患者复查可以用100mAs。 Purpose: To analyze the effect of various tube current settings(mAs) and its clinical value for adult cranial CT protocol. Materials and Methods: 110 patients, subdivided into three groups randomly, underwent a cranial non-helical CT scanning with 330-mAs, 130-mAs and 100-mAs, respectively but otherwise the same parameters. Two neuroradiologists visually scored the resulting images for quality in a blinded method and Mann-Whitney U test was used for comparisons. Representative regions of interest were chosen on selective reference levels and image noise, eonspicuity and the contrast-to-noise ratio were acquired. Statistical analyses were done by using Independent-Sample T test. Results: Subjective image scores of visibility of sylvian fissures were higher than that of internal capsules and grey-white differentiation in the cerebellum in 130-mAs and 100-mAs groups. The image scores of the same structures were significantly different between groups (P 〈 0.05 ). The noises of white matter in the cerebellum were higher than that of putamen and centrum semiovale in every group ( P 〈 0.05 ). The noises of the same level increased ( P 〈 0.05 ) and the CNR decreased ( P 〈 0.05 ) while conspicuity changed little ( P 〈 0.05 ) with the decrease of mAs. Compared to that of 330-mAs, T DLP of 130-mAs and 100-mAs decreased 60% and 70%, respectively. Conclusion: 130-mAs head CT images can be applied for adult and 100mAs used for follow-up examination.
出处 《中国医学影像学杂志》 CSCD 北大核心 2009年第5期336-339,共4页 Chinese Journal of Medical Imaging
关键词 CT 低辐射剂量 颅脑 CT radiation dosage head
  • 相关文献

参考文献10

  • 1Cohnen M,Fischer H,Hamacher J,et al.CT of the head by use of reduced current and kilovoltage:relationship between image quality and dose reduction.Am J Neuroradiol,2000,21(9):1654.
  • 2Gündodu S,Mahmutyaziciolu K,Ozdemir H,et al.Assessment of image quality of a standard and three dose-reducing protocols in adult cranial CT.Eur Radiol,2005,15(9):1959.
  • 3Mullins ME,Lev MH,Bove P,et al.Comparison of image quality between conventional and low-dose nonenhanced head CT.AJNR,2004,25(4):533.
  • 4Kalra MK,Maher MM,Toth TL,et al.Strategies for CT radiation dose optimization.Radiology,2004,230(3):619.
  • 5Christoph S,Chen XY.Dose optimization in pediatric CT:current technology and future innovations.Pediatr Radiol,2002,32(10):729.
  • 6Kalra MK,Prasad S,Saini S,et al.Clinical comparison of standard-dose and 50% reduced-dose abdominal CT:effect on image quality.Am J Roentgenol,2002,179(5):1101.
  • 7Kamel RI,Hemandez JR,Martin EJ,et al.Radiation dose reduction in CT of pediatric pelvis.Radiology,1994,190(3):683.
  • 8Britten AJ,Crotty M,Kiremidjian,et al.The addition of computer simulated noise to investigate radiation dose and image quality in images with spatial correlation of statistical noise:an example application to X-ray CT of the brain.The British J of Radiology,2004,77(916):323.
  • 9赖爱平,龚方戚,章士正.儿童CT检查射线剂量优化原则[J].中国儿童保健杂志,2005,13(4):342-343. 被引量:24
  • 10Chan CY,Wong YC,Chau LF,et al.Radiation dose reduction in paediatric cranial CT.Pediatr Radiol,1999,29(10):770.

二级参考文献17

  • 1METTLER FA JR, WIEST PW, LOCKEN JA, et al. CT scanning: patterns of use and dose[J]. J Radiol Prot, 2000,20:353-359.
  • 2ROEBUCK DJ. Risk and benefit in pediatric radiology[J]. Pediatr Radiol, 1999,29: 637-640.
  • 3ERIC J. Lessons we have learned from our children: cancer risks from diagnostic radiology[ J ]. Pediatr Radiol, 2002,32:700-706.
  • 4DIANNA DC, DONNA MM, KERRY TK. Strategies for formulating appropriate MDCT techniques when imaging the chest,abdomen, and pelvis in pediatric patients[J]. AJR, 2004,182:849-859.
  • 5KHURSHEED A, HILLIER MC, SHRIMPTION PC, et al.Influence of patient age on normalized effective doses calculated for CT examinations[J]. The British Journal of Radiology,2002,75: 819-830.
  • 6DAVID JE, CARL DE, ERIC JH,et al. Estimated risks of radiation induced fatal cancer from pediatric CT[J]. AJR, 2001,176:289-296.
  • 7WALTER HUDA. Dose and image quality in CT[J]. Pediatr Radiol, 2002,32: 709-713.
  • 8WARE DE, HUDA W, MERGO PJ, et al. Radiation effective doses to patients undergoing abdominal CT examinations[J].Radiology, 2001,210: 645-650.
  • 9HUDA W, SCALZETTI EM, ROSKOPF M. Effective doses to patients undergoing thoracic computed tomography examinations[J]. Med Phys, 2000,27: 838-844.
  • 10HUDA W, CHAMBERLAIN CC, ROSENBAUM AE, et al.Radiation doses to infants and adults undergoing head CT examinations[J]. Med Phys, 2001,28: 393-399.

共引文献23

同被引文献184

引证文献23

二级引证文献117

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部