摘要 采用自行研制的往复摩擦磨损试验机,研究了不同成分人工滑液润滑的实验条件下UHMWPE/Ti_6Al_4V关节材料的生物摩擦学性能,并利用扫描电子显微镜观察其磨损表面形貌.结果表明,不同成分人工滑液润滑条件下的UHMWPE/Ti_6Al_4V摩擦副的摩擦系数以及UHMWPE试样的体积磨损量有明显差异:去离子水润滑性很差,因其润滑的UHMWPE/Ti_6Al_4V摩擦副的摩擦系数以及UHMWPE试样的体积磨损量都最大;Hank's液的减磨性虽然较好,但是润滑性很差;小牛血清去离子水溶液的浓度在20%~40%(V/V)时,UHMWPE/Ti_6Al_4V摩擦副的摩擦系数以及UHMWPE试样的体积磨损量都很小. Biotribological behavior of artificial joint material of UHMWPE/Ti6 Al4 V was investigated under lubricating condition with different components of some artificial synovia by a homemade reciprocating friction and wear tester, and topographies of worn surfaces were observed by SEM. The results showed that friction coefficient of UHMWPE/Ti6Al4V rubbing pairs and volume loss of wear of UHMWPE were clearly different under lubricating condition with different components of some artificial synovial;' lu- bricating behavior of deionized water was not good because friction coefficient of UHMWPE/Ti6Al4 V rubbing pairs and volume loss of wear of UHMWPE were both maximal; wear-reducing behavior of Hank's solution was better, but its lubricating behavior was worse; friction coefficient of UHMWPE/Ti6 Al4 V rubbing pairs and volume loss of wear of UHMWPE were both little under lubricating condition with calf serum deionized water whose content range of calf serum was between 20% and 40% (V/V).
出处 《华中师范大学学报(自然科学版)》 CAS CSCD 北大核心 2009年第3期444-447,共4页 Journal of Central China Normal University:Natural Sciences
基金 国家自然科学基金资助项目(50575164 50775164) 湖北省教育厅科技项目(Q200718004) 湖北省中青年面上基金(2007ABA372)
关键词 超高分子量聚乙烯(UHMWPE) 钛合金(Ti6Al4V) 生物摩擦学性能 人工滑液 ultrahigh molecular weight polyethylene (UHMWPE) titanium alloy(Ti6 Al4 V) biotribological behavior artificial synovia
  • 相关文献

参考文献9

  • 1Cooper J R,Dowson D,Fisher J.Birefringent studies of polyethylene wear specimens and acetabular cups[J].Wear,1991,151:391-402.
  • 2Dowson.D,Harding.R T.The wear characteristics of ultrahigh molecular weight polyethylene against a high density alumina ceramic under wet (distilled water) and dry condition[J].Wear,1982,75:313-331.
  • 3Dowson D.A comparative study of the performance of metallic and ceramic femoral head components in total replacement hip joints[J].Wear,1995,190:171-183.
  • 4Barrett T S,Stachowiak G E,Batchelor A W.Effect of roughness and sliding speed on the wear and friction of ultrahigh molecular weight polyethylene[J].Wear,1992,153:331-350.
  • 5Weightman B,Light D.The effect of the surface finish of alumina and stainless steel on the wear rate of UHMW-polyethylene[J],Biomaterials,1986,7:20-24.
  • 6韩可芳.生命现象中的力学模型[J].华中师范大学学报(自然科学版),1989,23(4):503-506. 被引量:1
  • 7John H D.Tribology of natural and artificial joints[M].New York:Elsevier Scientific Publishing Company.1981:123-173.
  • 8Dowsoa D,Wright V.An introduction to the bio-mechanics of joints and joint replacement[M].London:Mechanical Engineering Publication Ltd,1981:120-145.
  • 9Hills B A.Boundary lubrioation in vivo[J].Journal of Engineering in Medcine,2000,214(1):83-90.