期刊文献+

无线多通道表面肌电信号采集系统设计 被引量:31

Design of Wireless multi-channel surface EMG acquisition system
下载PDF
导出
摘要 本文设计了一种无线多通道表面肌电信号(surface electromyography,SEMG)采集系统,该系统包括多通道的无线传感器和信号接收部分。传感器可独立的穿戴于人体表面,以线形差分电极获取表面肌电信号,对其进行放大、滤波、A/D变换,并用无线的方式按本文设计的通信协议发送给接收部分。接收部分对各传感器的数据进行整合,并通过USB接口传输给电脑进行存储、显示和处理。每个传感器体积为35mm×20mm×11mm,重量仅13g(含电池),一次充电可工作9个小时,无线通信距离达7.5m,采集到的信号噪声低于?70dB(肌电信号1mV代表0dB)。该设计大大提高了电极安放的便利性,采集设备的便携性与人体的安全性,且避免了工频干扰,能够满足基于表面肌电信号的手势或姿势识别等研究的要求。 A wireless multi-channel surface electromyography (sEMG) acquisition system consisting of a number of wireless sensors and receivers is introduced in this paper. Each wireless sensor contains linear differential surface EMG electrodes, amplifier, filter, A/D converter and 2.4 GHz transceiver, with the size of 35 mm × 20 mm × 11 mm and the weight of just 13 grams (including battery). The sensors are worn on the surface of the body independently to collect sEMG signals caused by human movement, and a battery charge can support 9 hours of collection. The receiver receivs the sEMG data from multiple sensors synchronously and which is transmitfed to computer through USB for storing, displaying and processing. The wireless communications range reaches up to 7.5 meters, and the noise signal collected is less than -70 dB. The design of our wireless system can improve greatly the convenience and security of sensor placement, reduce the power-frequency interference in sEMG signals, and satisfy the basic requirement of the data acquisition in the research of sEMG -based human movement recognition.
出处 《电子测量与仪器学报》 CSCD 2009年第11期30-35,共6页 Journal of Electronic Measurement and Instrumentation
基金 国家"863"计划(编号:2009AA01Z322)资助项目 国家自然科学基金(编号:60703069)资助项目
关键词 表面肌电信号 肌电信号采集 2.4GHz无线通信 surface EMG sEMG acquisition 2.4GHz wireless communication
  • 相关文献

参考文献8

  • 1DELUCA C J. Physiology and mathematics of myoelectric signals [J]. IEEE Transactions on Biomedicine Engineering, 1979, 26 (6): 313-325.
  • 2ZHAO Z Y, CHEN X, ZHANG X, et al. Study on online gesture sEMG recognition[C]. LNCS, 2007, 4681: 1257- 1265.
  • 3FUKUDA O, TSUJI T. A human-assisting manipulator teleo- perated by EMG signals and arm motion[J]. IEEE Transactions on Robotics and Automation, 2003, 19(2): 210-222.
  • 4ASO S, SASAKI A, HASHIMOTO H, et al. Driving electric car by using EMG interface[C]. IEEE International Conferences on Cybernetics Intelligent Systems, 2006: 1-5.
  • 5朱昊,辛长宇,吉小军,施文康.表面肌电信号前端处理电路与采集系统设计[J].测控技术,2008,27(3):37-39. 被引量:20
  • 6TOSHIYUKI K, OSAMU A. Proposal of anticipatory pattern recognition for EMG prosthetic hand control[C]. IEEE International Conferences on SMC, 2008: 897- 902.
  • 7钱晓进,杨基海,冯焕清,蒋鑫.肌电检测中消除工频干扰的方法[J].中国医疗器械杂志,2003,27(4):260-263. 被引量:9
  • 8HERMIE J. Development of recommendations for SEMG sensors and sensor placement procedures[J]. Journal of Electromyography and Kinesiology. 2000, 10(5): 361-374.

二级参考文献10

  • 1杨福生.独立变量分析及其在生物医学工程中的应用[A]..南京:99'中国生物医学电子学学术年会论文集[C].,.34-37.
  • 2Dylsys Incorporated Surface Electromyography:Detection and Recording.http://www delysy.com 1996.
  • 3Basmajian JV et al.Muscles Alive 5th ed Baltimore Williams&Wilkins Co.1985.
  • 4Hyvannen A. Independent component analysis: A Tutonal. http://www.cis.hut.fi/project/ICA.
  • 5童诗白.模拟电子技术基础[M].北京:高等教育出版社,2000..
  • 6EZ-USB FX2 technical reference manual version 2.0 [ EB/OL]. Cypress Semiconductor Corp data book ,2001.
  • 7Waiter Oney. Programming the Microsoft Windows driver model [ M ]. Microsoft, 1999.
  • 8钱峰.EZ-USBFX2单片机原理,编程及应用[M].北京:北京航空航天大学出版社,2006.
  • 9聂东,张红兵.USB2.0规范与EZ-USB FX2高速外设控制器[J].国外电子元器件,2002(7):37-41. 被引量:3