摘要
目的评价64层CT冠状动脉成像中前瞻性轴面扫描不同kV设置对射线剂量及图像质量的影响及临床应用价值,探讨〈1mSv冠状动脉检查的可行性。方法从拟行冠状动脉成像的患者中连续选取90例体质量指数(BMI)〈21.5kg/m2,且心率满足前瞻性门控轴面扫描条件的患者,平均心率(54±6)次/min(bpm)。90例患者数字表法随机分成A、B2组,A组45例扫描时管电压为100kV,B组45例扫描时管电压为120kV,根据患者BMI选择管电流,记录扫描中患者所受射线剂量,并对其图像质量进行评价(5分优,1分差)。采用配对t检验对2组的射线剂量和图像质量进行统计学分析。结果A组平均扫描剂量为(0.88±0.32)mSv,图像质量评分平均为(4.32±0.28)分;B组平均扫描剂量为(1.47±0.41)mSv,图像质量评分平均为(4.52±0.34)分。A、B组之间图像质量评分差异无统计学意义(t=0.771,P〉0.05):A组扫描剂量与B组相比较差异有统计学意义(t=12.125,P〈0.0t),平均减少40%。结论64层CT冠状动脉成像前瞻性轴面扫描中,对于BMI〈21.5kg/m2。的患者,可采用100kV管电压扫描,其辐射剂量可降到1mSv以下,并且其图像质量能满足临床诊断需要。
Objective To evaluate radiation dose and image quality of prospective electrocardiograph (ECG)-triggered coronary 64-slice CT angiography (CTA) with different kV, and the feasibility of coronary scan with 〈 1 mSv radiation dose. Methods Ninety consecutive patients[ heart rate: (54 ±6) bpm, BMI 〈 21.5 kg/m2 ] underwent 64-slice CTA using prospective ECG-triggered axial scan protocol. They were divided into two groups: 45 were in group A( 100 kV) , 45 were in group B( 120 kV) , and their electric currents were tailored to patient's BMI. The radiation dose was recorded and the image quality(5 fur excellent, 1 for bad)was blindly evaluated. Statistical t-test was performed between two groups on dose and image quality. Results The radiation dose of group A, B was ( 0. 88 ±0. 32 ) , ( 1.47 ±0. 41 ) mSv respectively. There was no significant difference between image quality of group A with 100 kV (4. 32±0. 28) and that of group B with 120 kV (4. 52 ±0. 34) (t =0. 771 ,P 〉0. 05). And radiation dose of the patient using axial scan protocol at 100 kV was decreased by 40% compared with that of group B at 120 kV(t = 12. 125, P 〈 0. 01 ). Conclusion Prospective ECG-triggered axial scan in 64-slice coronary CTA can significantly reduce radiation exposure. For low-BMl ( 〈 21.5 kg/m2 ) patients, super low dose scan ( 〈 1 mSv) at 100 kV using prospective ECG-triggered axial scan could be selected. And the image quality can meet clinical diagnostic needs.
出处
《中华放射学杂志》
CAS
CSCD
北大核心
2009年第12期1255-1259,共5页
Chinese Journal of Radiology