磁力驱动轴流血泵溶血实验和动物实验 被引量:2

Hemolysis and animal experiments with the magnetically drived axial flow ventricular assist device
下载PDF
导出
摘要 目的对研制开发的一种新型的磁力外驱动轴流式心室辅助血泵的血液相容性能进行测试。方法利用特制血袋作为模拟循环管道,羊血作为循环介质,采用标准溶血指数衡量体外溶血实验性能。通过3只山羊12h在体实验衡量其在体适应性。结果实验测得轴流血泵体外实验标准溶血指数(NIH)为(0.158±0.043)mg/L。3例实验动物12h在体辅助无机械故障,血泵辅助后实验动物血液中游离血红蛋白(FHb)开始上升,最高达到164.8mg/L。结论磁力外驱动轴流血泵实验结果比较理想,值得进一步改进。 Objective To design and develop a magnetically drived axial left ventricular assist device,and evaluate its performance. Methods For the test of hemolysis in vitro mock circuit,a particular blood bag was used as a mock circulation loop and the sheep blood were used as the circulating medium. The normalized index of hemoglobin (NIH) was used to measure the hemolysis. The in vivo adaptability was tested through 12 hours in animal experiments with 3 goats. Results The NIH for the axial blood pump was (0.158 ± 0.043) mg/L. There were no mechanical breakdowns in 12 hours animal experiments; the free hemoglobin (FHb) of each experimental animal started rising just after circulation; the highest value was 164.8 mg/L. Conclusion The results obtained from the in vivo testing and animal experiment indicated that this is an acceptable design for blood pump,and worth to further improvement.
出处 《生物医学工程与临床》 CAS 2010年第1期1-4,共4页 Biomedical Engineering and Clinical Medicine
基金 北京市青年科研资助项目(QN2008-002) 国家高技术研究发展计划(863计划)(2006AAO2Z4E8)
关键词 轴流血泵 溶血实验 动物实验 axial blood pump hemolysis testing animal testing
  • 相关文献

参考文献9

  • 1Sayer G, Naka Y, Jorde UP. Ventricular assist device therapy [J]. Cardiovasc Ther, 2009, 27(2): 140-150.
  • 2] Xiao XJ, Luo ZX, Ye CX, et al. The short-term pulsatile ventricular assist device for postcardiotomy cardiogenic shock: a clinical trial in China[J]. Artif Organs, 2009, 33(4): 373- 377.
  • 3Recently-Approved Devices. http://www.fda.gov/MedicalDevices/Products and Medical Procedures/Device Approvals and Clearances/Recently-Approved Devices.htm.
  • 4蔺嫦燕,孙衍庆,董培青,魏文宁,王建华,王秀敏.微型血泵的研制及其模拟实验研究[J].中国生物医学工程学报,1997,16(1):64-69. 被引量:18
  • 5钱坤喜,曾培,茹伟民,袁海宇,封志刚,王芳群,李岚.临床用叶轮式心室辅助装置的改进与完善[J].中国生物医学工程学报,2000,19(4):461-463. 被引量:6
  • 6肖学钧,范瑞新,成安衡,罗征祥,张镜芳,吴若彬,丁以群,高婉梅,叶椿秀.新型国产隔膜气动左心辅助泵(罗叶泵)的临床试用[J].岭南心血管病杂志,2002,8(1):43-45. 被引量:13
  • 7Mizuguchi K, Damm G, Benkowsky R, et al. Development of an axial flow ventricular assist device: in vitro and in vivo evaluation[J]. Artif Organs, 1995, 19(7): 653-659.
  • 8Song X, Throckmorton AL, Untaroiu A, et al. Axial flow blood pumps[J]. ASAIO J, 2003, 49(4): 355-364.
  • 9蔺嫦燕,李冰一,陈力争,姜以岭,王景,孙衍庆,侯晓彤.Ⅱ型叶片轴流血泵的研制及体外流量实验[J].中国生物医学工程学报,2002,21(2):149-153. 被引量:10

二级参考文献29

  • 1钱坤喜,顾洪熙,林匡定,费青.非脉动流低溶血叶轮血泵[J].中国生物医学工程学报,1989,8(4):258-264. 被引量:7
  • 2蔺嫦燕,孙衍庆,董培青,魏文宁,王建华,王秀敏.微型血泵的研制及其模拟实验研究[J].中国生物医学工程学报,1997,16(1):64-69. 被引量:18
  • 3赫尔姆特 舒尔茨,泵原理、计算与结构,1991年,106页
  • 4李文斌,磁力应用工程,1991年
  • 5钱坤喜,中国生物医学工程学报,1990年,9卷,2期
  • 6洛尔金,离心泵与轴流泵,1978年,83页
  • 7Mitamora Y. The valve-pump an axial non-pulsatile blood pump. Trans Am Soc Artif Internal Organs[J]. 1991,37:M520~522.
  • 8Macris MP, Myers TJ, Jarvik R, et al. In vivo evaluation of an intra-ventricular electric axial flow pump for left ventricular assistance[J]. Trans Am Soc Artif Internal Organs 1994,40:M719~722.
  • 9Kenji Yamazaki, Toshio Mori, Jun Tomioka, et al. The cool seal system: a practical solution to the shaft seal problem and heat related complications with implantable rotary blood pumps[J]. Trans Am Soc Artif Internal Organs,1997,43:M567~571.
  • 10Longya Xu, Fengxiang Wang, Minghua Fu, et al. Analysis of a new PA motor design for a rotary dynamic blood pump[J]. Trans Am Soc Artif Internal Organs 1997,43:M559~564.

共引文献34

同被引文献36

  • 1吴广辉.轴流式心脏辅助装置及其发展现状[J].北京生物医学工程,2004,23(4):306-308. 被引量:5
  • 2徐晓佳,李帆.轴流式血泵的研究进展[J].北京生物医学工程,2005,24(5):394-397. 被引量:5
  • 3MH Kwon,Abbas Ardehali,JD Moriguchi,et al.Use ofventricular assist device as a bridge to cardiac transplantation[J].Texas Heart Institute Journal,2009,36(3):214-216.
  • 4Menon AK.Current success and risks of mechanical circulatory supportU].Anasthesiol Intensivmed Notfallmed Schmerzther,2011,46(6):434-441.
  • 5Birks EJ.Left ventricular assist devices[J].Heart.Jan,2010,96(1):63-71.
  • 6Gopalan RS,Arabia FA,Noel P,et al.Hemolysis from aortic regurgitation mimicking pump thrombosis in a patient with a heaMate II left ventricular assist device: a case report[J]aSAIO J,2012,58(3):278-280.
  • 7Luckraz H,Woods M,Large SR,et al.And hemolysis goes on entricular assist device in combination with veno-venous lemofiltration[J].Ann Thorac Surg,2002,73(2):546-548.
  • 8Fossttm TW,Morley D,Olsen DB,et al. Complications common to ventricular assist device support are rare with 90 days of DeBakey VAD support in calveslJ].ASAIO J,2001,47(3):288-292.
  • 9ASTM F1841-97(2005)standard practice for assessment of hemolysis in continuous flow blood pumps[DB/OL][2012-10- 02].Http://www.astm.org/Standards/F1841.htm.
  • 10Nishinaka T,Schima H,Roethy W,et al.The DuraHeart VAD,a magnetically levitated centrifugal pump:the University of Vienna bridge-to-transplant experience[J].Circ J,2006,70(11):1421-1425.

引证文献2

二级引证文献17

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部