期刊文献+

过硫酸铵消化-砷铈催化分光光度法结合控温油浴测定尿碘的结果观察 被引量:1

Measurement of urinary iodine using ammonium persulfate digestion-As-Ce catalytic spectrophotometry with temperature-controlled oil bath
原文传递
导出
摘要 目的观察尿样经控温油浴消化后尿碘结果的精确度和准确度。方法控温油浴代替电热消解仪消解尿样,采用"尿中碘的砷铈催化分光光度测定方法"(WS/T107-2006)测定尿碘。结果标准曲线线性范围为0~300μg/L,相关系数平均为-0.9998。精确度:测定尿碘为(67.6±10)、(110.5±10)、(223.4±10)μg/L的3份尿样,变异系数分别为3.1%、2.3%、4.7%(n=6)。准确度:测定碘含量为(78.4±9)、(150.5±10)μg/L的2份尿碘国家标准物质,相对标准偏差分别为4.8%和2.8%(n=6)。结论控温油浴可以取代电热消解仪消解尿样进行尿碘测定,具有控温精准,制作容易的优点。 Objective To determine the precision and accuracy of measurement of urinary iodine levels in urine samples digested by temperature-controlled oil bath. Methods The urinary iodine level in the urine samples, digested by temperature-controlled oil bath instead of electrothermal digestion devices, was determined according to the As-Ce catalytic spectrophotometry method for measurement of urinary iodine ( WS/T 107 - 2006 ). Results The linear range of the standard curve was 0 - 300 μg/L, the average correlation coefficient - 0. 9998. In terms of the precision, the coefficients of variation in determining the urinary iodine levels of three samples with the contents of (67.6±10), (110.5 ± 10), (223.4± 10) μg/L were 3.1%, 2.3%, and 4.7% (n =6), respectively. In terms of the accuracy, the relative standard deviations in determining the levels of two national standard substances of iodine-containing urine samples with the iodine content of (78.4±+ 9) and ( 150.5 ± 10)μg/L were 4.8 % and 2.8 % ( n = 6), respectively. Conclusion Temperature-controlled oil bath, featuring precise temperature control and ease of making, is a good substitute for electrothermal devices for the measurement of urinary iodine levels.
出处 《疾病监测》 CAS 2010年第1期72-73,78,共3页 Disease Surveillance
关键词 尿 分光光度计 恒温水浴箱 控温油浴箱 urine iodine spectrophotometer constant temperature water bath temperature-controlled oil bath tank
  • 相关文献

参考文献4

二级参考文献9

  • 1阎玉芹,张亚平,刘列钧,刘嘉玉,李卫东,华基礼,陈祖培.尿中碘的过硫酸铵消化—砷铈催化分光光度测定方法[J].中国地方病学杂志,2004,23(6):582-585. 被引量:134
  • 2Pino S, Fang SL, Braverman LE. Ammunium persulfate:a new and safe method for measuring urinary iodine by ammonium persulfate oxidation. Exp Clin Endocrinol Diabetes, 1998,106(suppl 3):22-27.
  • 3Sullivan KM, May S, Maberly GM. Urinary Iodine assessment:a manual on survey and laboratory methods. 2nd. Atlanta (USA):PAMM/UNICEF, 2000. 35-45.
  • 4ICCIDD, UNICEF, WHO. Assessment of iodine deficiency disorders and monitoring their elimination: a guide for programme managers.Geneva,WHO/NHD, 2001, 73-75.
  • 5Suldin OP. Controversies in urinary iodine determinations. Clin Biochem, 2002, 35:575-579.
  • 6Allouch S, Lejeune PJ, Barbaria J, et al. Urinary iodine analysis:an alternative method for digestion of urine samples. Clin Chem,2004, 50:780-782.
  • 7WS/T107-1999.尿碘的砷-铈催化分光光度测定方法[S].
  • 8阎玉芹,陈祖培.对标准化尿碘测定方法应用过程中几个问题的讨论与建议[J].中国地方病学杂志,1999,18(4):298-300. 被引量:50
  • 9魏开效,吕荣春,谢宜凤.尿碘外质控样检测分析探讨[J].中国卫生检验杂志,2002,12(2):234-235. 被引量:4

共引文献154

同被引文献7

  • 1Blow N. Lab automation: tales along the road to automation E J 1. Nat Methods,2008,5 ( 1 ) : 109 - 112.
  • 2Wurdemann HA, Aminzadeh V, Dai JS, et al. Category-based food ordering processes[ J 1. Trends Food Sci Technol,2011,22(1) :14 -20.
  • 3McGrath TF, Elliott CT, Fodey TL. Biosensors for the analysis of microbiological and chemical contaminants in food I J ]- And Bioanal Chem,2012,403 ( 1 ) :75 - 92.
  • 4Li W, Cui ZG, Kan B, et al. China Lab based molecular subtyping network for bacterial disease surveillance-PulseNet China ~ J 1. Disease Surveillance, 2010 26 ( 1 ) : 1 - 4. ( in Chinese).
  • 5Meldrum D. Automation for genomics, part one: preparation for sequencing E J ]. Genome Res ,2000 ,10 ( 8 ) : 1081 - 1092.
  • 6Clarke SC, Diggle MA. Automated PCR/sequence template purification[ J]. Mol Biotechnol,2002,21 (3) :221 -224.
  • 7Greub G, Prod' horn G. Automation in clinical bacteriology: what system to choose? [ J]. Clin Microbiol Infect,2011,17 (5) :655 -660.