期刊文献+

基于化学发光磁酶免疫的HCG检测系统研究 被引量:10

Development of HCG detecting system based on chemiluminescent magnetic enzyme-linked immunoassay
下载PDF
导出
摘要 介绍了一种基于化学发光磁酶免疫分析技术测定人绒毛膜促性腺激素(HCG)的检测系统。该检测系统通过良好的屏蔽和抗干扰设计,以高速高频的FPGA器件完成光子计数,实现对HCG的定量检测。对HCG抗原溶液的检测结果表明,检测系统在HCG抗原浓度为0.06~600ng/mL范围内线性关系良好,相关系数R=0.951。对检测系统进行重复性测试,变异系数为0.61%,重复性较好。该系统具有小型便携、灵敏度高、重复性好等特点,在疾病标志物现场检测领域具有较为广阔的应用前景。 An HCG detecting system based on chemiluminescent magnetic enzyme-linked immunoassay is described.The system could detect weak chemiluminescence sensitively using photon-counting method with a special photo-electronic detection unit.In order to improve the detecting precision,FPGA is used to count the photon pulse signal.HCG sample with concentration of 0.06~600 ng/mL was measured and the correlation coefficient between HCG concentrations is 0.951.The variation coefficient of the system is 0.61%.With high sensitivity and high precision,the system will have great potential in trace disease marker measurement.
出处 《仪器仪表学报》 EI CAS CSCD 北大核心 2010年第2期312-317,共6页 Chinese Journal of Scientific Instrument
基金 国家高技术研究发展计划(2007AA03Z428) 国家自然科学基金(60801032) 中科院电子所创新项目(08SQCX-9151)资助
关键词 人绒毛膜促性腺激素 光子计数 化学发光磁酶免疫 human chorionicgonadotropin photon-counting chemiluminescent magnetic enzyme-linked immunoassay
  • 相关文献

参考文献14

  • 1RODA A, PASINI P, MIRASOLI M, et al. Biotechnological applications of bioluminescence and chemiluminescence[J]. Trends Biotechnol., 2004, 22:295-303.
  • 2WU A H B. A selected history and future of immunoassay development and applications in clinical chemistry[J]. Clin Chim Acta, 2006, 369:119.
  • 3司曙光,杨国栋.真空紫外光电倍增管测试技术研究[J].光电技术,2007,48(1):32-36. 被引量:1
  • 4MASSA J S, BOLLER G S, WALKER A C, et al. Time-of-flight optical ranging system based on time-correlated single-photon-counting[J]. Applied Optics,1998,37(31): 7298-7304.
  • 5VIDALI M, BARI M, FONTANAROSA D, et al. Development of a flexible MAPMT photon-counting read-system[J]. Nuclear Instruments and Methods in Physics Research A, 2005, 524: 375-382.
  • 6HAMAMATSU. Photomultiplier tubes basics and applications[M]. 3rd edition, 2006: 126-133.
  • 7BRONSTEIN I, EDWARDS B, VOYTA J C. 1,2- Dioxetanes: Novel chemiluminescent enzyme substrates. applications to immunoassays[J]. Journal of Bioluminescence and Chemiluminescence, 1989, 4:99-111.
  • 8TSUKAGOSHI K, JINNO N, NAKAJIMA R. Development of a micro total analysis system incorporating chemiluminescence detection and application to detection of cancer markers[J]. Anal. Chem., 2005, 77(6): 1684-1688.
  • 9石根,唐宝军,王栩,赵利霞,林金明.微孔板化学发光酶免疫分析法分析人血清中游离前列腺特异性抗原[J].分析化学,2007,35(11):1541-1547. 被引量:15
  • 10徐晔,张微,毛敏.基于ATP生物化学发光法微生物含量检测仪[J].仪器仪表学报,2007,28(2):367-371. 被引量:5

二级参考文献60

共引文献53

同被引文献113

引证文献10

二级引证文献23

相关作者