期刊文献+

牙髓活力对根管长度测量仪准确性影响的临床研究 被引量:3

Clinical study of accuracy of working length determination by Root ZX
下载PDF
导出
摘要 目的观察牙髓状态对Root ZX根尖定位仪测定根管长度的准确性有无影响。方法收集临床须行根管治疗的患牙94例,共156个根管,使用Root ZX确定工作长度,预备根管后拍主尖片验证工作长度的测定是否准确,统计分析活髓牙组和死髓牙组准确率有无差异。结果在活髓牙和死髓牙的根管预备中,利用Root ZX测定根管工作长度的准确率差异无统计学意义(P>0.05)。结论牙髓活力对根管长度测量仪准确性无显著影响。 Objective To observe the accuracy of working length determination by Root ZX in the teeth with different pulp vitality.Methods 94 cases requiring root canal therapy(156 root canals)were prepared and the working length(WL) was determined by Root ZX.Master gutta percha points were inserted into the canals and radiographs were taken to observe whether the WL determination was accurate.And the statistical analysis was performed.Results The difference of accuracy in WL determination by Root ZX between the teeth with different pulp vitality was not significant(P0.05).Conclusion The accuracy of Root ZX working in different pulp vitality is not adversely affected.
作者 王靖
出处 《重庆医学》 CAS CSCD 北大核心 2010年第10期1233-1234,共2页 Chongqing medicine
关键词 根管长度 根管测量 根管治疗 working length electronic apical locator endodontic treatment
  • 相关文献

参考文献12

二级参考文献59

共引文献77

同被引文献41

  • 1张杨,庄最新,张爱华.根管扩大程度影响根管长度电测仪准确性的研究[J].广东牙病防治,2007,15(1):9-11. 被引量:3
  • 2范兵,边专,樊明文.牙体牙髓临床治疗Ⅱ. 根管预备的工作长度、宽度和锥度[J].中华口腔医学杂志,2006,41(3):184-187. 被引量:27
  • 3刘思逸,梅陵宣.根管长度电测法的发展史及现状[J].国际口腔医学杂志,2006,33(5):361-363. 被引量:13
  • 4周学东,岳松龄,主编.牙体牙髓痛治疗学.北京:人民卫生出版社、2004.188-189.
  • 5Yu DC,Schilder H. Cleaning and shaping the apical third of a root canal system. Gen Dent,2001,49(3):266-270.
  • 6Martinez-Lozan MA,Fomer-Navarro L,Sanchez-Cortes JL,et al. Methodological considerations in the determination of working length. Int End,2001,34:371-376.
  • 7Kuttler. Microscopic investigation of root apexes. AmDent Assoc,1955, 50: 544-552.
  • 8Dunlap CA,Remeikis NA,BeGole EA. An in vivo evaluation of an electronic apex locator that uses the ratio method invital and necrotic canals. J Endod,1998,24(1):48-50.
  • 9Shabahang S,Goon WW,Gluskin AH. An in vivo evaluation of Root ZX electronic apex locator. Endodon,1996,22:616-618.
  • 10Ebrahim AK,Wadachi R,Suda H. Ex vivo evaluation of the ability of four different electronic apex locators to determine the working length in teeth with various foramen diameters. Australian Dental Journal, 2006,51 (3): 258-262.