摘要
用偏最小二乘法(PLS)校正了火焰原子吸收分析中Eu324.753nm对Cu324.754nm吸收线重叠干扰,用铜空心阴极灯同时测定了合成样及稀土氧化物中的Eu和Cu。
Partial least squares regression was used to compensate for spectral 'overlap' interference of Eu 324 753 nm with Cu 324 754 nm in atomic absorption spectrometry.We could only use the copper element hollow cathode lamp to simultaneous determine Eu and Cu in synthetic samples and rare earth oxide,and obtained satisfactory results.
出处
《光谱学与光谱分析》
SCIE
EI
CAS
CSCD
北大核心
1999年第1期84-86,共3页
Spectroscopy and Spectral Analysis
基金
江西省自然科学基金
关键词
火焰原子吸收
PLS
稀土氧化物
铕
铜
测定
Flame atomic absorption, Partial least squares regression, Rare earth oxide, Europium, Copper