摘要 A scattering correction method for a panel detector based cone beam computed tomography system is presented. First,the x-ray spectrum of the system is acquired by using the Monte Carlo simulation method.Secondly,scattered photon distribution is calculated and stored as correction matrixes by using the Monte Carlo simulation method according to scanned objects and computed tomography system specialties.Thirdly,scattered photons are removed from projection data by correction matrixes.A comparison of reconstruction image between before and after scattering correction demonstrates that the scattering correction method is effective for the panel detector based cone beam computed tomography system. A scattering correction method for a panel detector based cone beam computed tomography system is presented. First, the x-ray spectrum of the system is acquired by using the Monte Carlo simulation method. Secondly, scattered photon distribution is calculated and stored as correction matrixes by using the Monte Carlo simulation method according to scanned objects and computed tomography system specialties. Thirdly, scattered photons are removed from projection data by correction matrixes. A comparison of reconstruction image between before and after scattering correction demonstrates that the scattering correction method is effective for the panel detector based cone beam computed tomography system.
出处 《Chinese Physics B》 SCIE EI CAS CSCD 2010年第8期609-613,共5页 中国物理B(英文版)
基金 Project supported by the National Natural Science Foundation of China(Grant Nos.60672104 and 10527003) the National Basic Research Program of China(Grant No.2006CB705705) the Joint Research Foundation of Beijing Education Committee, China(Grant No.JD100010607)
关键词 平板探测器 CT系统 散射光子 校正方法 锥束 蒙特卡罗模拟方法 校正矩阵 X射线谱 scattering correction, Monte Carlo simulation, panel detector, cone beam computed tomography system
  • 相关文献

参考文献16

  • 1Rinkel J, Gerfault L, Esteve F and Dinten J M 2007 Phys. Med. Biol. 52 4633.
  • 2Siewerdsen J and Jeffray D 2001 Med. Phys. 28 220.
  • 3Zhu L, Bennett N R and Fahrig R 2006 IEEE Transactions on Medical Imaging 25 1573.
  • 4Ning R and Tang X 2004 Med. Phys. 31 1195.
  • 5Neitzel U 1992 Med. Phys. 19 475.
  • 6Lo J Y, Floyd C E, Baker J A and Ravin C E 1994 Med. Phys. 21 435.
  • 7Naimuddin S, Hasegawa B and Mistretta C A 1987 Med. Phys. 14 330.
  • 8Seibert J A and Boone J M 1988 Med. Phys. 15 567.
  • 9Wiegert J, Bertram M, Rose G and Aach T 2005 Proc. SPIE 5T45 271.
  • 10Jarry G, Graham S A and Moseley D J 2006 Med. Phys. 33 4320.