校准ICP-MS检出限的测量不确定度评定 被引量:6

Evaluation of Measurement Uncertainty of Detection Limits of ICP-MS by Calibrating
下载PDF
导出
摘要 依据JJF1159-2006《四级杆电感耦合等离子体质谱仪(ICP-MS)校准规范》,利用国家规定的标准溶液GBW(E)130242(Be,In,Bi混合标液),校准了ICP-MS的检出限并对其不确定度进行了评定。 Based on JJF1159-2006 "Calibration Procedures of Quadrupole Inductively Coupled Plasma Mass Spectromete(ICP-MS) ",the detection limits of ICP-MS are calibrated using the state standard solution GBW(E) 130242 containing Be,In and Bi elements,and the measurement uncertainty is evaluated.
作者 施江焕
出处 《计测技术》 2010年第4期47-49,共3页 Metrology & Measurement Technology
关键词 ICP-MS 检出限 测量不确定度 ICP-MS detection limit measurement uncertainty
  • 相关文献

参考文献4

二级参考文献75

  • 1蒋杰,方力,张鹤颖,窦文华.无线传感器网络最小连通覆盖集问题求解算法[J].软件学报,2006,17(2):175-184. 被引量:90
  • 2Grenville Holland. Plasma Source Mass Spec-trometry : New Developments and Applications. Cambridge : Royal Society of Chemistry,1999,299.
  • 3Grenville Holland,Scott D Tanner. Plasma Source Mass Spectrometry: the New Millennium. Cambridge : Royal Society of Chemistry, 2001,427.
  • 4Howard E Taylor. Inductively Coupled Plasma-Mass Spectrometry : Practices and Techniques,San Diego : Academic Press, 2001, 294.
  • 5Akbar Montaser. Inductively Coupled Plasma Mass Spectrometry,New York : J. Wiley, 1998,964.
  • 6Roy W Morrow, Jeffrey S Crain. Applications of Inductively Coupled Plasma-Mass Spectrometry to Radionuclide Determinations. West Conshohocken, Pennsylvania : ASTM, c1998, 145.
  • 7Roy W Morrow, Jeffrey S Crain. Applications of Inductively Coupled Plasma-Mass Spectrometry to Radionuclide Determinations. West Conshohocken, Pennsylvania: ASTM, c1995, 145.
  • 8Wangzhao Zhu. Advanced Inductively Coupled Plasma Mass Spectrometry Analysis of Rare Earth Elements. Environmental Applications,Rotterdam : Balkema, 1999,157.
  • 9Narong Praphairaksit, R S Houk. Attenuation of Matrix Effects in Inductively Coupled Plasma Mass Spectrometry with a Supplemental Electron Source inside the Skimmer. Analytical Chemistry, 2000, 72(11): 2351~2355.
  • 10Scott D Tanner, Lisa M Cousins, Douglas DJ. Reduction of Space Charge Effects Using a Three-Aperture Gas Dynamic Vacuum Interface for Inductively Coupled Plasma-Mass Spectrometry. Appl. Spec., 1994,48 (11).