GENSCREENULTRA HIV抗原抗体酶联免疫诊断试剂盒的敏感性评价 被引量:1

Sensitivity evaluation of GENSCREEN~ULTRA HIV Ag-Ab kit
原文传递
导出
摘要 目的评价BIO-RAD公司开发研制的人类免疫缺陷病毒(HIV1+2)抗原抗体酶联免疫诊断试剂盒(GENSCREEN~ ULTRA HIVAg-Ab),对不同来源样本检测的敏感性。方法以法国生物梅里埃公司生产的HIV1+2抗原/抗体ELISA检测试剂为参比试剂,分别使用评价试剂和参比试剂检测621例样本,分析两种试剂检测HIV抗原/抗体的敏感性。结果评价试剂和参比试剂检测502份HIV感染者/AIDS患者临床血浆标本均为阳性,符合率为100%,敏感性为100%。评价试剂和参比试剂检测92份HIV分离培养上清液,评价试剂检出89份阳性、参比试剂检出31份阳性,表明评价试剂检测HIV-1p24抗原的敏感性高于参比试剂。另外,又检测了27份HIV抗原系列稀释样本,计算出评价试剂检测HIV抗原的敏感性是参比试剂的4~8倍。结论评价试剂检测HIV抗体的效果与参比试剂相同,检测HIV-1p24抗原的敏感性高于参比试剂。 Objective To evaluate the sensitivity of GENSCREEN ULTRA HIV Ag-Ab kit made by BIO-RAD for detecting different samples.Methods Vironostika HIV Uni-Form Ⅱ Ag/Ab was chosen as reference which was made by Biomerieux.621 samples were tested with prepared kit and reference kit.And then,the difference in sensitivities to detect HIV Ag/Ab between the prepared kit and the reference kit was analyzed.Results The 502 serum samples taken from HIV-infected people or AIDS patients were all HIV positive by using the prepared kit and the reference kit,with a consistence rate of 100% and sensitivity of 100%.To test HIV P24 antigen 92 HIV culture supernatant samples were detected by using the two kits,of which 89 samples were positive by the prepared kit and 31 samples were positive by the reference kit,indicating that the sensitivity of the prepared kit was higher than that of the reference kit in detecting HIV Ag.Additionally,by testing HIV P24 antigen with 27 serial diluted samples it was calculated that the sensitivity of the prepared kit was 4-8 times higher than that of the reference kit in detecting HIV Ag accurately.Conclusions Compared with the reference kit,the prepared kit has the same effect in detecting HIV Ab and higher sensitivity in detecting HIV Ag.
出处 《中国艾滋病性病》 CAS 2010年第4期341-343,共3页 Chinese Journal of Aids & STD
关键词 人类免疫缺陷病毒 抗体 抗原 GENSCREENULTRA HIV Ag-Ab诊断试剂盒 敏感性 HIV Ab Ag GENSCREEN ULTRA HIV Ag-Ab Kit Sensitivity
  • 相关文献

参考文献7

  • 1Schreiber GB, Bush MP, Kleinman SH, et al. The risk of transfusion-transmitted viral infection[J]. The Retrovirus Epidemiology Donor Study. N EnglJ Med, 1996, 334(26): 1685-1690.
  • 2Whyte GS, Savoia HF. The risk of transmitting HCV, HBV or HIV by blood transfusion in Vietoria[J]. Med J Aust, 1997, 166 (11):584- 586.
  • 3Weber B, Mbargane Fall E H, Berger A-M, et al. Reduction of diagnostic window by new fourth-generation human immunodeficiency virus screening assays[J]. J Clin Microbiol, 1998, 36:2235 -2239.
  • 4Gurtler L, Muhlbacher A, Michl U, et al. Reduction of the diag nostie window with a new combined p24 antigen and human immu nodeficiency virus antibody screening assay[J]. J Virol Methods 1998, 75:27-38.
  • 5Ly TD, Laperche S, Brennan C, et al. Evaluation of the sensitivity and specificity of six HIV combined p24 antigen and antibody assays[J]. J VirolMethods, 2004, 122:185-194.
  • 6Weber B, Meier T, Enders M, et al. Fourth generation human immunodeficiency virus(HIV) screening for p24 antigen close the second diagnostic window in primary HIV infection[J]. J Clin Virol, 2002, 25:357-359.
  • 7Meier T, Knoll E, Henkes M, et al. Evidence for a diagnostic window in fourth generation assays for HIV[J]. J Clin Virol, 2001, 23:113-116.

同被引文献10

引证文献1

二级引证文献5

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部