γ-谷氨酰基转移酶参考方法的建立与性能评估

Establishment and evaluation of the reference methods for gamma-glutamyltransferase
下载PDF
导出
摘要 目的建立γ-谷氨酰基转移酶(GGT)参考方法,对其性能进行评价,为临床常规检测GGT提供可比性和溯源性。方法根据国际临床化学联合会(IFCC)颁布的GGT一级参考测量程序(37℃)要求,建立GGT一级参考方法(其中包括所需仪器、试剂、样本的准备与校准等实验条件的试验),并对其进行精密度、准确性和最大线性范围检测。结果 GGT参考方法的精密度评价A、B 2个水平样本变异系数(CV)总均<1%,符合精密度性能要求。有证参考物质(ERM-AD452/IFCC)均值为113.27 U/L,均值与靶值的偏倚为-0.64%,在允许范围内。GGT的最大线性范围为413.81 U/L,高于IFCC公布的GGT参考方法的最大线性范围(276.40 U/L)。结论 GGT参考方法已基本建立,精密度、准确度及线性范围等性能指标均符合方法要求,为我国参考实验室网络的建立提供了基础,临床应用将更加广泛。 Objective To establish and evaluate the reference method of gamma-glutamyltransferase(GGT) and provide the comparability and traceability of determining GGT routinely in clinical laboratory.Methods The reference method for GGT including the preparation and calibration of instruments,reagents and samples was established according to the documents issued by International Federation of Clinical Chemistry and Laboratory Medicine(IFCC).The precision,accuracy and linear range of the method were detected.Results According to the accuracy of GGT reference method,A and B horizontal samples were evaluated [total coefficient of variation(CVt)1%].For the reference material(ERM-AD452/IFCC),the average value was 113.27 U/L,and the bias between the average value and the target value was-0.64% within the allowance range.The upper limit of the linear range of GGT was 413.81 U/L,which was higher than the data published by IFCC(276.40 U/L).Conclusions The reference method for GGT has been established basically.The analytic performance such as precision,accuracy and linear range conforms to the requirements.The reference method has been established,which can be used widely in clinical laboratory routinely and for research.
出处 《检验医学》 CAS 北大核心 2010年第9期693-696,共4页 Laboratory Medicine
关键词 Γ-谷氨酰基转移酶 参考方法 有证参考物质 血清酶学 Gamma-glutamyltransferase Reference method Certified reference material Serum enzyme
  • 相关文献

参考文献2

  • 1Schumann G,Bonora R,Ceriotti F,et al.IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C.International Federation of Clinical Chemistry and Laboratory Medicine.Part 6.Reference procedure for the measurement of catalytic concentration of gamma-glutamyltransferase[J].Clin Chem Lab Med,2002,40(7):734-738.
  • 2王涛,齐丽丽,刘海燕,田真,何振坤,李淑葵,但万春,徐国宾.六种临床诊断酶学测定参考方法的应用及国际比对结果分析[J].中华检验医学杂志,2008,31(3):264-269. 被引量:15

二级参考文献8

  • 1符颖操,罗茜.实验室间比对结果分析统计方法的探讨[J].理化检验(物理分册),2006,42(6):295-299. 被引量:55
  • 2徐国宾,吴南,王清涛.要重视血清酶学测定的标准化工作[J].临床检验杂志,2007,25(3):161-164. 被引量:25
  • 3Schumann G, Bonom R, Ceriotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C. Part 4.Reference procedure for the measurement of catalytic concentration of alanine aminotransferase. Clin Chem Lab Med, 2002, 40:718-724.
  • 4Schumann G, Bonora R, Ceriotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C. Part 2. Reference procedure for the measurement of catalytic concentration of creatine kinase. Clin Chem Lab Med, 2002, 40:635-642.
  • 5Schumann G, Bonora R, Ceriotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C. Part 3. Reference procedure for the measurement of catalytic concentration of lactate dehydrogenase. Clin Chem Lab Med, 2002, 40:643-648.
  • 6Schumann G, Aoki R, Ferrero C A, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C. Part 8. Reference procedure for the measurement of catalytic concentration of alpha- amylase. Clin Chem Lab Med, 2006, 44-1146-1155.
  • 7Schumann G, Bonora R, Ceriotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C. Part 6. Reference procedure for the easurement of catalytic concentration of gammaglutamyltransferase. Clin Chem Lab Med, 2002, 40:734-738.
  • 8Schumann G, Bonora R, Ceriotti F, et al. IFCC primary reference procedures for the measurement of catalytic activity concentrations of enzymes at 37 degrees C. Part 5. Reference procedure for the measurement of catalytic concentration of aspartate aminotransferase. Clin Chem Lab Med, 2002, 40:725- 733.

共引文献14

相关作者

内容加载中请稍等...

相关机构

内容加载中请稍等...

相关主题

内容加载中请稍等...

浏览历史

内容加载中请稍等...
;
使用帮助 返回顶部